Chapter 10
235
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
[:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO
Syntax
[:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO {ON|OFF|1|0}
[:SENSe]:CORRection:COLLect:STAN3:RANGe:AUTO?
Description
Sets on/off of auto ranging when the load correction is executed. When this function is set
at off, the range which is selected by
[:SENSe][:FIMPedance]:RANGe[:UPPer]
is used for
the load correction.
For auto ranging when a measurement is executed, use
[:SENSe][:FIMPedance]:RANGe:AUTO
NOTE
The functionality of this command is same as
[:SENSe]:CORRection:COLLect:LOAD:RANGe:AUTO
Parameter
Query response
{1|0}<newline><^END>
Related Commands
[:SENSe]:CORRection:COLLect[:ACQuire] on page 233
[:SENSe][:FIMPedance]:RANGe[:UPPer] on page 262
:SYSTem:PRESet on page 279
Equivalent key
sequence
[
Meas Setup
] - CORRECTION - LOAD RNG- AUTO/FIX.
Description
ON or 1 (initial
setting)
Measures the load standard with auto range.
OFF or 0
Measures the load standard measurement range which has been set
just before the load standard measurement is performed (auto
range function is invalid).