Chapter 10
261
Command Reference
E4981A SCPI Commands
10
. Comm
and
Refe
ren
ce
[:SENSe][:FIMPedance]:RANGe:AUTO
Syntax
[:SENSe][:FIMPedance]:RANGe:AUTO {ON|OFF|1|0}
[:SENSe][:FIMPedance]:RANGe:AUTO?
Description
Selects the measurement range mode from the auto range (automatic range switching) or
hold range (fixed range).
Setting the measurement range (set with the
[:SENSe][:FIMPedance]:RANGe[:UPPer]
command) automatically selects the hold range mode.
Parameters
Query response
{1|0}<newline><^END>
Related commands
[:SENSe][:FIMPedance]:RANGe[:UPPer] on page 262
Equivalent key
sequence
[
Meas Setup
] - RANGE - AUTO|HOLD
Description
ON or 1 (initial value)
Specifies the auto range mode.
OFF or 0