Mask Testing
15
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
231
On Error
The
On Error
setting specifies the action(s) to take when the input waveform
does not conform to the mask. This setting supersedes the
Run Until
setting.
• Stop
— The oscilloscope will stop when the first error is detected (on the first
waveform that does not conform to the mask). This setting supersedes the
Minimum # of Tests
and
Minimum Time
settings.
• Save
— The oscilloscope saves the screen image when an error is detected. In
the Save Menu (press
[Save/Recall] > Save
), select an image format (*.bmp
or *.png), destination (on a USB storage device), and file name (which can be
auto-incrementing). If errors occur too frequently and the oscilloscope
spends all its time saving images, press the
[Stop]
key to stop acquisitions.
— The oscilloscope prints the screen image when an error is detected.
This option is only available when a printer is connected as described in
print the oscilloscope's display"
• Measure
— Measurements (and measurement statistics if your oscilloscope
supports them) run only on waveforms that contain a mask violation.
Measurements are not affected by passing waveforms. This mode is not
available when the acquisition mode is set to Averaging.
Note that you can choose to
or
Save
, but you cannot select both at the
same time. All other actions may be selected at the same time. For example, you
can select both
Stop
and
Measure
to cause the oscilloscope to measure and
stop on the first error..
You can also output a signal on the rear panel TRIG OUT BNC connector when
there is a mask test failure. See
"Setting the Rear Panel TRIG OUT
Source Lock
When you turn on Source Lock using the
Source Lock
softkey, the mask is
redrawn to match the source whenever you move the waveform. For example, if
you change the horizontal timebase or the vertical gain the mask is redrawn with
the new settings.
When you turn off Source Lock, the mask is not redrawn when horizontal or
vertical settings are changed.
Source
If you change the Source channel, the mask is not erased. It is re-scaled to the
vertical gain and offset settings of the channel to which it is assigned. To create
a new mask for the selected source channel, go back up in the menu hierarchy;
then, press
Automask
, and press
Create Mask
.
The Source softkey in the Mask Setup Menu is the same as the Source softkey in
the Automask Menu.
Test All
When enabled, all displayed analog channels are included in the mask test.
When disabled, just the selected source channel is included in the test.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...