296
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
22
Reference
CSV and ASCII XY file structure
In CSV or ASCII XY format the
Length
control selects the number of points per
segment. All segments are contained in the CSV file or in each ASCII XY data file.
For example: If the Length control is set to 1000 points, there will be 1000 points
(rows in the spreadsheet) per segment. When saving all segments there are three
header rows, so the data for the first segment starts at row 4. The second
segment's data starts at row 1004. The time column shows the time since the
trigger on the first segment. The top row shows the selected number of points per
segment.
BIN files are a more efficient data transfer format than CSV or ASCII XY. Use this
file format for fastest data transfer.
Minimum and Maximum Values in CSV Files
If you are running a Minimum or Maximum measurement, the minimum and
maximum values shown in the measurement display may not appear in the CSV
file.
Explanation:
When the oscilloscope's sample rate is 4 GSa/s, a sample will be taken every
250 ps. If the horizontal scale is set to 10 us/div, there will be 100 us of data
displayed (because there are ten divisions across the screen). To find the total
number of samples the oscilloscope will take:
100 us x 4 GSa/s = 400K samples
The oscilloscope is required to display those 400K samples using 640 pixel
columns. The oscilloscope will decimate the 400K samples to 640 pixel columns,
and this decimation keeps track of the min and max values of all the points that
are represented by any given column. Those min and max values will be displayed
in that screen column.
A similar process is used to reduce the acquired data to produce a record usable
for various analysis needs such as measurements and CSV data. This analysis
record (or
measurement record
) is much larger than 640 and may in fact contain
up to 65536 points. Still, once the # of acquired points > 65536, some form of
decimation is required. The decimator used to produce a CSV record is configured
to provide a best-estimate of all the samples that each point in the record
represents. Therefore, the min and max values may not appear in the CSV file.
Summary of Contents for DSO-X 2002A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 22: ...22 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 62: ...62 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 70: ...70 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 84: ...84 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Page 106: ...106 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 112: ...112 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 146: ...146 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 154: ...154 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 180: ...180 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 216: ...216 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 228: ...228 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 246: ...246 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 266: ...266 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 20 Utility Settings...
Page 352: ...352 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide Index...