N9069C Noise Figure Measurement Guide 79
Measurement Related Tasks and Concepts
Setting the Signal Analyzer to Make a Measurement
Configuring Fixed Loss Compensation
To configure fixed loss compensation follow the example below:
Step
Notes
1.
Press
MEAS SETUP
,
Loss Comp
tab.
2.
Press
Loss, Before DUT
or
After DUT
,
Loss
Comp Mode
,
Fixed
.
3.
Press
Fixed Loss
and enter the loss
compensation value and press the
dB
termination key.
The lower limit is –100.000 dB, the upper limit is
100.000 dB, and the default is 0.000 dB.
4.
Press
Temperature
, and use the numeric
keys or the knob to enter the temperature of
the device where the loss is occurring. This
will normally be room temperature, which is
290 K.
It is important that you enter the correct temperature. This
will give you better accuracy for you measurements.