Example 1. Monitoring LED Forward diode current 

(continued)

Step 7.  As a inal step we will check the residual current of the test ixture.
 

(1) Close the lid of the test ixture.

 

(2) Disconnect the Vs cable from the ixture.

 

(3) Check the current transient in the Roll view display and note the 

 

  inal residual current.

 

Note that the residual current quickly reaches the pA level and eventually 

 

stabilizes at the fA level (3 fA in about 100 seconds in this example).

Example 2. LED forward diode I/V characteristics

This example outlines how to plot LED current versus voltage (I/V) characteristics and it only applies to the B2985A/87A.

Note: If using the B2981A/83A then you need to plot the current measured by the B2981A/83A and the voltage applied by the external voltage source on a 
PC to make this same type of I/V plot.

The following steps outline how to plot the I/V characteristics of an LED.

Step 1.  (1) Press the [Meter View] function key at the bottom of the screen.
 

(2) Then press the [More ... 1 of 3] assist key twice so that the third set [More ... 3 of 3] of softkey labels is displayed on the  

 

 

assist keys at the right of the screen.

 

(3) Press the [Show VS Func.] assist key to show the VS function menu.

Step 2.  (1) Move the cursor by rotating the [Knob] until the cursor is on the VS Function and then select it by pushing the knob.
 

 

The VS Function is now in edit mode and the assist softkey menu at the right of the screen represents the available VS  

 

 Functions.

 

(2) Press on the [LINEAR SINGLE] assist key.

(3) press             key.

3.

(1) Press                key, then 

(2) press              key two times until             key appears.

1.

2.

(1) Click on the VS Function by         , and the VS 

menu options are displayed.

(2) Press           assist key.

1.

Move the cursor on the VS Function by rotating          .

2.

Summary of Contents for B2981A

Page 1: ...ght Technologies Low Current Semiconductor Measurements Using the B2980A Series Ammeter Keysight B2981A B2983A Femto Picoammeter Keysight B2985A B2987A Electrometer High Resistance Meter Technical Ove...

Page 2: ...lexibility 3 The ability to improve data quality through averaging iltering and zero cancelation 4 Integrated capability to analyze measurement data via histograms time domain graphs etc The Keysight...

Page 3: ...rovide stable measurements and a lexible coniguration for low current measurement applications Floating DUT coniguration Figure 1 shows a circuit block diagram using the loating ammeter with a DC volt...

Page 4: ...ovides the following beneits The ability to synchronize the voltage source output with the measurement timing The ability to display the voltage and current relationship in an X Y graph on the instrum...

Page 5: ...at its source or to ilter out the noise from the data Of course if the noise is symmetrically distributed then you can simply use the average of the histogram Figure 4 shows an example of this proced...

Page 6: ...using the B2981A B2983A Femto Picoammeter requires some modiications to the software key sequences and steps shown in this example although the ammeter setup is identical Of course when using the B29...

Page 7: ...then follow the steps shown below to set the output voltage to zero volts Step 3 1 Press the Voltage Source On Off key to enable the voltage source to output 0 V Note If using the B2981A 83A then set...

Page 8: ...nge the voltage to approximately 1 0 V Step 6 1 Continue rotating the Knob and increasing the voltage until the current is between 10 to 20 mA Note If using the B2981A 83A then increase the external v...

Page 9: ...the B2981A 83A and the voltage applied by the external voltage source on a PC to make this same type of I V plot The following steps outline how to plot the I V characteristics of an LED Step 1 1 Pre...

Page 10: ...s the Show Trigger assist key The trigger sub panel is now displayed If the Trigger mode is set to MANUAL as shown in step 5 then perform the following steps 5 Using the Knob place the cursor over the...

Page 11: ...ance and stable picoammeter that can measure down to 100 aA with 10 aA effective resolution The B2980A Series employs a loating ammeter design which provides considerable freedom to use the ammeter in...

Page 12: ...rs www keysight com find channelpartners Get the best of both worlds Keysight s measurement expertise and product breadth combined with channel partner convenience www keysight com find precisionMEASU...