![Keysight Technologies B2960A Demo Manual Download Page 9](http://html1.mh-extra.com/html/keysight-technologies/b2960a/b2960a_demo-manual_1962179009.webp)
09 | Keysight | Using an External Trigger to Generate Pulses with the B2960A – Demo Guide
14) Or press to set it to
INFINITY
if you’d like to continue to source pulses. If you
set the trigger count to INFINITY be sure you abort the process at the end of the test.
4. Disable an internal source wait time
The B2960A series has internal wait time parameters to adjust the timing to source and
measure in order to have the measurements made under the exact source condition as
specified. However, it is required to disable it to source pulses at the exact timing as
specified by external trigger signals
1) If you aren’t on the top of the Function menu, press repeatedly to return to
the top level.
2) Press , , and then press to open the
Wait Control
dialogue.
3) Press and select to disable the
Source wait time
.
On the top level of the Function menu
In the middle level of the Function menu
No tree structure can be seen
Some tree structure can be seen
(1) Press Config
(2) Press Common
(3) Press Wait