Keysight B1542A User’s Guide, Edition 10
5-27
PLSDIV Test Definitions
PLSDIV IV SMU
PLSDIV IV SMU
This test definition controls SMUs to apply the primary and secondary sweep
voltage, measures current, and displays the I-V characteristics.
Device Parameters
The following parameter is available to set the test condition.
IDispRange
Maximum value of the Y1 axis, in A. 1 mA to 100 mA, 6 digits
or 1 fA resolution.
Test Parameters
The following parameters are available to set the source output.
SMU1
Channel number of the SMU used for the primary sweep
source. Slot1 to Slot8. Maximum channel number depends on
the configuration of DC source monitor.
StartV1
Primary sweep start voltage.
-
100 to 100 V, 6 digits or 1
V
resolution.
StopV1
Primary sweep stop voltage.
-
100 to 100 V, 6 digits or 1
V
resolution.
StepV1
Primary sweep step voltage.
-
100 to 100 V, 6 digits or 1
V
resolution.
Comp1
Primary sweep current compliance, in A.
-
100 mA to 100 mA,
6 digits or 1 fA resolution.
•
StartV1, StopV1, StepV1
The parameters set the primary sweep source used to apply the drain voltage.
The number of sweep points is given by the following formula. Then ignore the
numbers after the decimal point. If StepV1=0, the measurement voltage is
StartV1 only.
Number of sweep points = (StopV1
-
StartV1) / 1
•
StartV2, StopV2, StepV2
The parameters set the secondary sweep source used to apply the gate voltage.
The number of sweep points is given by the following formula. Then ignore the
numbers after the decimal point. If StepV2=0, the measurement voltage is
StartV2 only.
Number of sweep points = (StopV2
-
StartV2) / 1
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......