Keysight B1542A User’s Guide, Edition 10
4-5
Performing Measurement
Before Measurement
Before Measurement
Confirm or perform the followings before starting measurement.
1. Performing system reset
Perform system reset. See
“Performing System Reset” on page 4-6
2. Deciding device delay time
This is needed to set the DeviceDelay value of the Device Parameters on the
pulsed IV test setup screen.
Connect the device under test (MOSFET) to the pulsed IV test system as shown
in
, and measure the delay time between the gate
monitor pulse and the drain monitor pulse by using the oscilloscope. Then
decide the device delay time shown in
.
3. DcSwitch value on Extended Setup dialog box of pulsed IV test setup
Set Yes to use the pulse/dc switch. Do not set No.
4. SWAddress value on Extended Setup dialog box of pulsed IV test setup
Set GPIB address of the switch controller.
Figure 4-3
Device Delay Time
After adjustment
Before adjustment
Drain monitor pulse
Device Delay
Gate monitor pulse
Drain monitor pulse
Gate monitor pulse
Summary of Contents for B1542A Pulsed IV Package
Page 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Page 9: ...1 Introduction ...
Page 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Page 19: ...2 Installation ...
Page 41: ...3 Performing System Setup and Compensation ...
Page 55: ...4 Performing Measurement ...
Page 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Page 69: ...5 PLSDIV Test Definitions ...
Page 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Page 109: ...6 PLSDIV TIS Commands ...
Page 133: ...7 Status Code and Error Messages ...
Page 159: ......