Arbitrary Block Program or Response Data
Definite-length block data <Block> allows any type of device-dependent data to be programmed or
returned as a series of 8-bit binary data bytes. This is particularly useful for transferring large
quantities of data or 8-bit extended ASCII codes.
Device Clear
Device Clear is an IEEE-488 low-level bus message that you can use to return the instrument to a
responsive state. Different programming languages and IEEE-488 interface cards provide access to
this capability through their own unique commands. The status registers, the error queue, and all
configuration states are left unchanged when a Device Clear message is received.
Device Clear performs the following actions:
If a measurement is in progress, it is aborted.
The instrument returns to the trigger idle state.
The instrument's input and output buffers are cleared.
The instrument is prepared to accept a new command string.
The ABORt command is the recommended method to terminate an instrument
operation.
Typical Command Processing Times
The table below documents some typical, average command processing times (milliseconds) for
various commands and queries. This can help you determine the impact of some common SCPI
commands on total test time.
The command processing time is the time until the next command is accepted. It does not include
hardware response time.
Command
GPIB
[1]
USB Description
*CLS
5
6 Clear the status data.
*RCL
233
230 Recall the contents of a state storage location.
*RST
233
230 Perform a device reset.
*SAV
13
14 Save the current settings.
FREQuency
15
17 Set the AC output frequency.
MEASure:CURRent:AC?
333
333 Query the AC current.
MEASure:VOLTage:AC?
333
333 Query the AC voltage.
OUTPut OFF
19
23 Turn the output OFF.
OUTPut ON
9
11 Turn the output on.
VOLTage
16
18 Set the AC voltage.
[1]
Using Keysight IO Libraries.
3 SCPI Programming Reference
100
Keysight AC6800B Series Operating and Service Guide
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...