235
Specifications and Regulatory Information
Specifications
Specifications
Each laser line is assumed to have a linewidth (including modulation sidebands) of
less than 10 GHz.
All specifications apply when the instrument is in the following modes:
• NORMAL update mode unless noted.
Refer to “Measurement rate” on page 50
.
• Configured to measure narrowband devices. Specifications do
not
apply when the in-
strument is configured to measure broadband devices.
band devices and chirped lasers” on page 46
Wavelength
Range
700-1650 nm (182-428 THz)
Absolute accuracy (lines separated by
≥
30 GHz)
±
3 ppm (
±
0.005 nm at 1550 nm,
±
0.004 nm at 1310
nm)
Differential accuracy
(characteristic)
±
2 ppm
Minimum resolvable separation
a
(characteristic)
20 GHz (0.16 nm at 1550 nm, 0.11 nm at 1300 nm)
Display resolution
0.001 nm, normal update mode
0.01 nm, fast update mode
a. Signals of equal amplitude. For lines that are separated by less than 30 GHz, wavelength accuracy is reduced.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......