CALCulate2 Subsystem
WLIMit:STOP:FREQuency
Sets the stopping frequency for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP:FREQuency {?|{<real> | MINimum | MAXimum }}
<
real
> is a frequency value that is within the following limits:
Constant
Description
MINimum
start wavelength limit
MAXimum
428.2750 THz
Attribute Summary
Non-sequential command
Preset State: 249.8271 THz
*RST State: 249.8271 THz
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wavelength limit. The default units for
the <
real
> parameter are Hz. The stop frequency value must be greater than or equal
to the start frequency value or the stop frequency will be clipped to the start fre-
quency and a “Data out of range” error will be generated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command
has the time to complete before sending any more commands to the instrument.
Refer to “Always force the Keysight 86120B to wait for non-sequential com-
mands” on page 92
for more information.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......