User’s Guide 85523-90005
1-9
855xxA/B Series CalPods and 85523B CalPod Controller
Usage Information
3.11 Select Install Software - Windows XP and 2000 from the list.
If there is a prompt to completely remove the selected application, perform the following:
This step
(Step 3.13)
only applies to a network analyzer with firmware version
A.09.
xx
.xx or older. Version A.
10
.4
9
.
11
or higher does not require enabling
option 301 or 302.
To use the CalPod Installation Package that was downloaded in
Installation Package for Windows XP and Win7 OS” on page 1-7
, the CalPod firmware license
option (301 or 302) has to be enabled using a keyword that is specific for your PNA model and
serial number. This keyword was sent to you by Keysight. Use the following instructions to enter
the keyword in the PNA using the PNA front panel HARDKEYs and [softkeys].
1. On the PNA, press SYSTEM > [Service] > [Option Enable].
2. In the Option Enable dialog box, use the pull-down menu to select the CalPod firmware
option you bought with your initial purchase. If you don’t know which option was purchased,
refer to the papers in the shipment container.
3. Enter the keyword.
4. Select Permanent > Enable.
5. Click Yes to restart the PNA and enable the new firmware.
3.12 Wait for an Install Shield Wizard screen with a “Finish” button. When it appears, select Finish.
3.13 Enable the CalPod firmware license (option 301 or 302).
1. Click Yes.
2. At the next dialog window, wait for approximately five seconds before clicking Finish.
3. Repeat Step 3.11 to reinstall the driver application.