5-16 Electronic Calibration Modules Reference Guide
Specifications and Characteristics
Characteristic Performance
5-
N4431D (7-16)
The characteristic performance in
the following table applies to N4431D (
if any of
the ports consist of 7-16 connector
). The data describes performance when
measuring “thru p
ath” A-B, C-D, A–D, A-D, B–C, and B-D.
N4431D (Type-N 50
Ω
)
The characteristic performance in the following table applies to N4431D
Option 020 (Type-N 50
Ω
female connectors on all ports). The data describes
performance when measuring “thru path” A-B, C-D, A–D, A-D, B–C, and B-D.
Table 5-28 Characteristic Performance for N4431D (m/f 7–16 connectors)
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz
5 GHz to 7.5 GHz
Transmission
Tracking (±dB)
b
0.18
0.05
0.08
Load Match (dB)
35
42
40
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance indicated in this
table.
b. Values based on using the network analyzer N5232B 400.
Table 5-29 Characteristic Performance for N4431D (Type-N 50
Ω
) Option 020
a
Parameter
Frequency Range
DC to 10 MHz
10 MHz to 5 GHz
5 GHz to 9 GHz
9 GHz to 13.5 GHz
Transmission
Tracking (±dB)
b
0.18
0.05
0.08
0.12
Load Match (dB)
35
43
41
38
a. When applied power exceeds –7 dBm, calibration results will be degraded from the performance indicated in this
table.