1
Laser Safety Labels
Laser class 1M label
Figure 2
Class 1M Safety Label - 81960A
A sheet of laser safety labels is included with the laser module as required.
In order to meet the requirements of IEC 60825-1 we recommend that you
stick the laser safety labels, in your language, onto a suitable location on
the outside of the instrument where they are clearly visible to anyone using
the instrument.
WARNING
Please pay attention to the following laser safety warning:
•
The laser output can be controlled by GUI, GPIB state command and
the front panel button.
•
Do not switch on the instrument when there is no termination to the
optical output connector, to the optical fiber or to the attached
device. The laser radiation can seriously damage your eyesight.
•
The use of optical instruments with this product will increase eye
hazard.
•
Refer servicing only to qualified and authorized personnel.
CAUTION
The built in laser diode is active whenever the instrument is powered on,
therefore disabling the output is not sufficient to establish eye safe
conditions.