Keysight 16860 Series Portable Logic Analyzer Service Guide
39
Testing 16860 Performance
3
15 Save the configuration by selecting
File
then
Save As …
.
16 In the
Save As
dialog, type the file name
MultipleClocks_NoMarker.ala
. For file options, select
Setup
Only
. Then click
Save
.
Create Configuration File 2 - Multiple Clocks (with Marker)
1 Start with the
MultipleClocks_NoMarker
configuration file. Select
File
then
Open
. In the Open dialog
select the file and then open. If asked about saving the current configuration, click
No
.
2 Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods. The Pod 1 leads should be connected to the generator as described
above.
3 From the main drop down menu, select
Setup
and then select
Bus/Signal.
The
Analyzer Setup
dialog
is displayed.
4 The activity indicators now show activity on the channels that are connected to the pulse
generator. Un-assign all channels. You can do this quickly by clicking on the left-most check
mark and dragging to the right across all of the other check marks.
5 Assign bits that are connected, Bits 2, 6, 10, 14.
Summary of Contents for 16860 Series
Page 1: ...Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 8: ...8 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 12: ...12 Keysight 16860 Series Portable Logic Analyzer Service Guide Contents...
Page 32: ...3 Testing 16860 Performance 32 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 34: ...3 Testing 16860 Performance 34 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 104: ...5 Troubleshooting 104 Keysight 16860 Series Portable Logic Analyzers Service Guide...
Page 136: ...7 Replaceable Parts 136 Keysight 16860 Series Portable Logic Analyzer Service Guide...
Page 138: ...Index 138 Keysight 16860 Series Portable Logic Analyzer Service Guide...