Functional
T
est
This
section
provides
the
functional
test
procedure
to
check
the
16454A
performance
.
The
functional
test
can
be
used
for
post
repair
function
verication.
Fixture
Impedance
Check
1.
P
erform
calibration
at
the
APC-7
R
terminal
of
the
measurement
instrument.
2.
Place
the
xture
(small)
on
the
calibrated
APC-7
R
terminal
of
the
measurement
instrument.
3.
Read
L
S
and
R
S
values
for
each
test
xture
.
The
guideline
is
as
follows:
T
able
5-2.
Fixture
Impedance
Check
Guideline
Fixture
Frequency
P
arameter
Guideline
Small
100
MHz
L
S
1
nH
6
0.5
nH
R
S
<
100
m
Large
100
MHz
L
S
5.5
nH
6
2.5
nH
R
S
<
300
m
Service
5-3