1. Safety 

1.1 Equipment markings

1.2 Operational safety

The Kewtech KT1700 is designed to be used by skilled persons in accordance with 
safe methods of work. If the KT1700 is used in a manner not specified by Kewtech, 
the protection provided by it may be impaired.

Inspect the KT1700 before using. If any damage is visible such as cracks in the 
casing the unit should not be used.

The intermittent beep and flashing blue LED indicate that the tester is functioning 
correctly.

The unit will power off automatically after 3 minutes to preserve battery life.

Marks listed below are used on this instrument. 

Caution - refer to the instruction manual

Instrument with double or reinforced insulation

This instrument is subject to WEEE Directive (2002/96/EC). Please 

contact our dealer near you for disposal.

Conforms to EU standards

CAT

IV

Measurement Category IV is applicable to test and measuring circuits 

connected at the source of the building’s low-voltage MAINS installation. 

This part of the installation is expected to have a minimum of one level of 

over-current protective devices between the transformer and connecting 

points of the measuring circuit.
Due to these high short-circuit currents (above 50kA) which can be 

followed by a high energy level, measurements made within these 

locations are extremely dangerous. Great precautions shall be made to 

avoid any chance of a short circuit.
Examples of CAT IV are measurements on devices installed before the 

main fuse or circuit breaker in the building installation.