Installation Instructions
Supreme Air LV Bench Hood
Fume Hood w/Auto-return Sash
Kewaunee Scientific Corporation • P.O. Box 1842 • Statesville, North Carolina 28687-1842 • Phone 704-873-7202
IILV05-SA 06/05-
3
Packing List
Preparation
1.
Place, level, and install the
Base Unit Assembly
that will support the
Fume Hood
as shown on page 4.
2.
Attach the
Fume Hood Worksurface
to the
Base Unit Assembly
at the corners using epoxy cement or similar adhesive.
Worksurface should be centered on units and overhang the front by 1/2". Worksurface must be level.
Shim as needed. Top should extend all the way to rear wall.
3.
Uncrate
Fume Hood
taking care to sort and inventory all parts.
4.
Remove
End Panels
from
Fume Hood
(by lifting up and out) and remove screws anchoring
Weight Box
to
Fume Hood
.
5.
Remove all packing from
Fume Hood
and any tape holding parts in place for shipping.
Fume Hood is packed fully assembled except for
Lower Airfoil
. The
Top Front Panel
, may be
removed (by lifting up and out) to fit through narrow doorways and/or for easier access.
Generally electrical fixtures and plumbing fittings are shipped installed, but may be shipped loose
depending on job requirements. Fittings may also be pre-wired and/or pre-piped. Fixtures are not
shown on assembly diagram as quantities and locations vary to meet individual requirements.
Installation
6.
Check that
Sash Belts
are properly routed over front pulleys, and engage rear pully sprockets.
7.
Check sash auto-sash return height and adjust as needed. (See Auto-return Adjustment Instructions on page 5)
Handle Frameless Sash with care - Laminated Safety Glass will crack if abused.
9.
Using as many workers as needed, lift assembled
Fume Hood
and place on leveled
Worksurface
being careful not to spread
bottom of hood. Make sure hood is pushed back as far as possible and centered on
Base Unit Assembly
. Check
End Panels
for
plumb and square.
10. Attach
Lower Airfow
to the
Worksurface
and
Fume Hood
using the included hardware as shown in Detail 1. Check that
Lower
Airfoil
is straight and level.
11. Coordinate connections of service fittings, electrical fixtures, and ductwork with the respective trades (i.e. mechanical, electrical,
HVAC). After connections are complete, replace interior
Access Panels
, if removed, and the exterior
End Panels
.
12. Check
Sash
again for level, smooth movement, and auto-return. Adjust as needed.
To level, loosen the two (2) set screws on one of the rear
Sash Pulleys
(see Figure 2), level sash, then retighten set screws.
13. Inspect the complete installation, remove any dust, dirt, or other debris. Check all moving parts for proper operation, and adjust
or lubricate as necessary.
Rear Baffles
may be removed for cleaning by removing the two Baffle StandoffKnobs and lifting up
and out on the baffle panels.
NOTICE
CAUTION
!
NOTICE
NOTICE
NOTICE
On Pallet
Fume Hood
Hardware Package (taped to side of hood)
Lower Airfoil (on top of hood)
Fume Hood Worksurface is packaged separately.
NOTICE
Hardware Package
(4-7)
10-24 x 3/4" Phillips Truss Head Machine Screw
see detail 1
with Nut and Flat Washer
(2)
#8 x 1/2" Sheet Metal Screws
(1)
Installation Instructions