Kett Riceter M Series Instruction Manual Download Page 2

1

First,confirm that the figures "8 TIMES 88.8%"

and several other marks appear when the

"POWER" button is pressed with your finger.

Should the figures not appear ,or      mark appear,

the batteries are exhausted. Replace all of the

batteries with new ones, according to the diagram

illustrated in the battery chamber.

2

Press the "SELECT" button and repeat until

the Selection Indicator of the sample you wish

to test appears at the bottom of the display.

3

Take one spoonful of the grain to be measured,

and place it evenly in one layer in the Sampling

Tray.

NOTE: Sort out unripe or degenerated grains from

the tray in order to avoid error.

4

Rotate the Crushing Handle counterclockwise,

and fully insert the Sampling Tray into the

Testing Chamber. 

Fully rotate the Crushing

Handle until stop is reached , so that the sample

may be well crushed. Press the "POWER " button,

the most reliable moisture content is then

displayed to be read directly without using any

conversion tables.

5

The following indicators each have  important

meanings. The  illuminator lights for 2 sec-

onds when the "POWER " button is pressed and

for 4 seconds when the moisture content is

displayed or the "SELECT" button is pressed.  The

Battery Indicator indicates the battery power

remaining.This indicator is displayed for approxi-

mately 2 seconds when the "POWER" button is

pressed.

INSTRUCTIONS

Battery Indicator

Full

Half

Empty

 Over Indicator

 Under Indicator

 Selection Indicator

The Over Indicator "    "is displayed when the

moisture content of the measured sample

exceeds the upper limit of the measurement

range,  or in the event of problems with the

measurement  operation ,or if irregularities are

detected. The Under Indicator "    " is displayed

when the moisture content of the sample is below

the lower limit of the measurement range.

NOTE: Over or under indicators may also be

displayed in the event that the measurement pro-

cedure is perfomed without a sample loaded on

the sampling tray, there is  condensation within

the testing chamber, or insulation is poor due to

high humidity or soiling.  If this should occur, thor-

oughly clean the testing  chamber and allow it to

dry through natural evaporation if damp.

6

The averaged moisture content of 2-9

sequential tests is indicated when the "AVE"

button is pressed with in 2 minutes after the tests.

In this case,"    " or  "    "marked tests are

disregarded. Numbers of tests being performed is

shown on the left part of the digital display. But in

the following cases, the reading of the number of

tests goes back to "1".

1. More than 2 minutes have passed after the

    tests.

2. After the  tests are already  averaged.

3. More than 9 tests are performed.

POWER OFF
1. Continue to press the "POWER" button for 2

seconds, the  unit will power off.

2. When inactive, the unit will automatically power

off affer 5 minutes.

After each measurement, clean the

Testing Chamber, Crushing Handle and

Sampling Tray with  the supplied brush.

Too Many

Insufficient

Evenly
Placed in
one layer