• If calibrating 4 points or fewer, calibrate only the number of calibration foils required and then after Step
8
,
select
ENT
on the display and press
key.
Step
Operation
Display
Explanation
7
,
or
,
Press
key and input the coating thickness of
calibration foil (100 µm) by
,
or
,
keys.
8
Press
key again, 100 µm thickness is fixed, and
the word on display is changing from Standard1 to
Standard2.
9
Repeat step
678
Measure and fix the
(300 µm) calibration foil
Measure and fix the
(500 µm) calibration foil
Measure and fix the
(700 µm) calibration foil
Proceed to the calibration of the second calibration foil
(300 µm). Confirm that the screen displays the words
Standard2 and repeat Step
678
.
Proceed to the calibration of the third calibration foil
(500 µm). Confirm that the screen displays the words
Standard3 and repeat Step
678
.
Proceed to the calibration of the fourth calibration foil
(700 µm). Confirm that the screen displays the words
Standard4 and repeat Step
678
.
····
····