4
DEFINITIONS
(continued)
cal
: (calibrate; yes or no) Select
YES
to calibrate the unit, select
no
to skip the calibration
procedure. Press the
↑
key to select YES or NO. Press the
E key to enter the displayed
selection.
cal LO
: (calibrate low) Apply an accurate 4 mA signal to the input and press the
E key. If the
calibration is successful the unit will display "
DONE cal lo
". If the calibration is not successful
the display will read "
DONE cal err
" see error codes page 7. Press the
M key to continue.
cal hi
: (calibrate high) Apply an accurate 20 mA signal to the input and press the
E key. If the
calibration is successful the unit will display "
DONE cal hi
". If the calibration is not successful
the display will read "
DONE cal err
" see error codes page 7. Press the
M key to continue.
loc code
: (lock code) Sets the 4 digit lock code to be entered when the unit prompts
ent code
.
This allows the user to gain access to the menu when the unit is locked. Press the
↑
key to
increment each digit. Press the
←
key to step to the next digit to the left. Press the
E key to
enter the displayed code. (Factory Default Code = 1000)
Record this number for later use!
loc unit
: (lock unit) Sets the panel lock ON or OFF. Press the
↑
key to select
yes
(ON) or
no
(OFF). Press the
E key to enter the displayed selection.
NOTE: A hardware jumper menu lockout is also available. (see Typical Wiring Pg. 2)
4