DEFINITIONS

(continued)

cal

: (calibrate; yes or no) Select 

YES 

to calibrate the unit, select 

no 

to skip the calibration

procedure.  Press the 

↑ 

key to select YES or NO.  Press the 

E key to enter the displayed

selection.

cal LO

: (calibrate low) Apply an accurate 4 mA signal to the input and press the 

E key.  If the

calibration is successful the unit will display "

DONE cal lo

".  If the calibration is not successful

the display will read "

DONE cal err

" see error codes page 7.  Press the 

M key to continue.

cal hi

: (calibrate high) Apply an accurate 20 mA signal to the input and press the 

E key.  If the

calibration is successful the unit will display "

DONE cal hi

".  If the calibration is not successful

the display will read "

DONE cal err

" see error codes page 7.  Press the 

M key to continue.

loc code

: (lock code) Sets the 4 digit lock code to be entered when the unit prompts 

ent code

.

This allows the user to gain access to the menu when the unit is locked.  Press the 

↑ 

key to

increment each digit.  Press the 

 key to step to the next digit to the left.  Press the 

E key to

enter the displayed code.   (Factory Default Code = 1000)

Record this number for later use!

loc unit

: (lock unit) Sets the panel lock ON or OFF.  Press the 

↑ 

key to select 

yes

 (ON) or 

no

(OFF).  Press the 

E key to enter the displayed selection.

NOTE:  A hardware jumper menu lockout is also available. (see Typical Wiring Pg. 2)

4

Summary of Contents for SQUIRT-R

Page 1: ...SQUIRT R LOOP POWERED INDICATOR INSTRUCTION MANUAL 99357 01 14 2021 http www kep com KESSLER ELLIS PRODUCTS 10 Industrial Way East Eatontown NJ 07724 800 631 2165 732 649 7100 Fax 732 649 7099...

Page 2: ...INDEX SPECIFICATIONS 1 DIMENSIONS 2 TYPICAL WIRING 2 DEFINITIONS 3 PROGRAMMING FLOWCHART 5 ERROR CODES 7 BATTERY REPLACEMENT 7 DECODING PART NUMBER 7 WARRANTY 7...

Page 3: ...Protected Over Current Protection to 60 mA 16 Bit resolution 1 sam ple every 2 seconds Low Cutoff supplied to inhibit indica tions at low flow rates CALIBRATION OPERATION Input Scaling Via front keypa...

Page 4: ...d Line Shows Outside Panel Dimension 4 00 Diameter 120 Panel Cutout 1 20 30 5 2 875 73 SQUIRTR 1 SQUIRTR 0 SQUIRTR 3NE SQUIRTR 2 SQUIRTR 3SS SQUIRTR 6 SQUIRTR 5 4 92 125 4 33 110 4 33 110 8 Screw Moun...

Page 5: ...ey to select the descriptor Press the E key to enter the selected descriptor rat desc ratemeter descriptor Sets the rate descriptor Choose rate per hour minutes seconds or blank Press the key to step...

Page 6: ...y If the calibration is successful the unit will display DONE cal hi If the calibration is not successful the display will read DONE cal err see error codes page 7 Press the M key to continue loc code...

Page 7: ...e to the next digit Press the E key to enter the displayed code If the code is correct display advances to clr tot if not display returns to run mode Press the key to step to the desired input type Pr...

Page 8: ...h individual digit of the low cutoff setting Press the key to advance to the next digit Press the E key to enter the displayed value Press the M key to skip and keep the existing value Press the key t...

Page 9: ...NEMA 4X Box SquirtR behind clear cover 3 Explosion Proof Housing 5 NEMA 4X Box SquirtR outside opaque cover Options ET Extended Temp 22 F to 158 F 30 C to 70 C H2 0 875 Hole for mounting styles 2 and...

Reviews: