3  Commissioning 

Compatible models: 
 

  KERN EMB (only models with weighing plate Ø 82mm) 

 

KERN EMB 200-3V 

3.1  Installation 

3.1.1  Prepare the weighing balance 

 

  If necessary, carry out necessary adjustment before installation of the density 

set.  

  Correct adjustment is no longer possible after the density kit has been 

installed.  

  For reasons of adjustment, take away the density set and place the standard 

weighing plate. 

 

  Disconnect scale from power supply. 

  Remove standard weighing plate. 

 

 

Summary of Contents for TYDB-04-A

Page 1: ...info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating and Installation Instructions Density determination set pour KERN EMB EMB V KERN YDB 04 Type TYDB...

Page 2: ...ontents 1 INTRODUCTION 3 1 1 SCOPE OF DELIVERY 3 2 DIMENSION MM 4 3 COMMISSIONING 5 3 1 INSTALLATION 5 3 1 1 How to prepare the weighing balance 5 3 1 2 Installing the density determination set 6 4 DE...

Page 3: ...ctions only describe the operation of the density determination set For further information on how to operate your balance please refer to the operating instructions supplied with each balance 1 1 Sco...

Page 4: ...4 YDB 04 BA_IA e 1910 2 Dimension mm Platform Hooks...

Page 5: ...are the weighing balance If necessary carry out necessary adjustment before installation of the density set Correct adjustment is no longer possible after the density kit has been installed For reason...

Page 6: ...ook into one of the two holes Place the platform for support of the glass beaker Arrange a beaker glass not included in the scope of delivery Bring liquid and instruments to the right temperature unti...

Page 7: ...nce calculates automatically the density and displays it For models without density calculation function the density must be calculated according to the following formula Density of sample A Weight of...

Page 8: ...r solids example water at 19 C SOLId is briefly displayed followed by the actual entered density of the aid liquid If necessary change as described later Enter the density of the aid liquid considerin...

Page 9: ...04 BA_IA e 1910 9 Density determination solids 7 Hang the sample on the hook 8 Press the weight of the sample in air is shown example 9 Wait for stability display 10 Immerse the sample into the aid li...

Page 10: ...ter s density example When connecting an optional printer press the result will be printed Printout example KERN YKB 01N D REF 0 9976 g cm 3 Density aid liquid D RSL 8 0409 g cm 3 Result density of th...

Page 11: ...eighing in air Hang the sample on the hook Wait for stability display Read off and record the weight value Weighing in liquid Immerse the sample into the aid liquid Wait for stability display Read off...

Page 12: ...the balance calculates automatically the density and displays it For models without density calculation function the density must be calculated according to the following formula Density of test liqui...

Page 13: ...bes will be described later While using the same sinker the entered density stays stored Skip the following steps for additional measurements and start by determining the density of the liquid step 7...

Page 14: ...4 BA_IA e 1910 Density determination liquids 7 Hang the sinker on the hook 8 Press the weight of the sinker in air is displayed example 9 Wait for stability display 10 Immerse the sinker into the test...

Page 15: ...result example When connecting an optional printer pressing of will print the result Printout example KERN YKB 01N D REF 8 0409 g cm 3 Sinker density D RSL 0 9984 g cm 3 Result Density of the test liq...

Page 16: ...ghing in air Hang the sinker on the hook Wait for stability display Read off and record the weight value Weighing in test liquid Immerse the sinker into the test liquid Wait for stability display Read...

Reviews: