Inadmissible disturbances from the environment may result in incorrect displays,
inaccurate measured values or incorrect behaviour of the medical device. The line
regulation is less than ±1kg unstable readings while measuring at rated weight
capacity.
Also in certain cases the baby scale MBA-M may cause such failures in other
devices. To eliminate problems of that kind, we recommend you to take one or
several of the measures listed below:
•
Change the alignment or distance of the device to the source of EMI.
•
Install or use the baby scale MBA-M on a place.
•
Connect the baby scale MBA-M to another power source.
•
For further questions please contact our customer services.
Disturbances may be caused by improper modification or add-ons to the device or
not recommended accessories (such as power units or connecting cables). The
manufacturer will not be responsible for these. Modifications may also result in a loss
of authorisation relating to the use of the device.
Devices emitting high frequency signals (mobile telephones, radio transmitters, radio
receivers) may cause interference in the medical device. For that reason do not use
them near the medical device. Chapter 6.4 contains details about recommended
minimum distances.