KERN SFB 100K-2HM Operating Instructions Manual Download Page 39

SFB-BA-e-1625 

39 

7.11  Animal weighing 

The animal weighing function is ideal for unstable loads.  
The weighing system calculates and displays a stable mean average from several 
weighing values. 
 
The animal weighing program may either be enabled by calling up menu block  

P3 OTH

“ or “

P4 OTH

 „

ANM

“ 

 „

ON

“ (See chap. 8) or by using the faster option of 

a key combination.  

 

 

 

The indicator shows 

HOLD 

as long as the animal weighing function remains enabled. 

 

 

 

 

  Place the load onto the weighing system and wait until it is fairly stable. 

  Press 

 and 

 at the same time, a signal sounds, meaning that the animal 

weighing function is enabled. 
During the calculation of a mean average you can add or remove loads as the 
mean average will be continuously updated.  

  To disable the animal weighing function press 

 and 

 at the same time. 

 

Summary of Contents for SFB 100K-2HM

Page 1: ...iegelei 1 D 72336 Balingen email info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating instructions Stainless steel scale KERN SFB Version 2 5 04 2016 G...

Page 2: ...fety Precautions 14 4 1 Pay attention to the instructions in the Operation Manual 14 4 2 Personnel training 14 5 Transportation Storage 14 5 1 Testing upon acceptance 14 5 2 Packaging return transport...

Page 3: ...Parts counting 38 7 11 Animal weighing 39 7 12 Lock keyboard 40 7 13 Display background illumination 40 7 14 Automatic switch off function AUTO OFF 41 8 Menu 42 8 1 Overview non verifiable models 43...

Page 4: ...10 minutes 30 minutes Stabilization time typical 2 sec Weighing unit kg Auto Off Available options Ambient temperature 10 C 40 C Moist environment 0 95 non condensing Electric Supply Input voltage 11...

Page 5: ...nutes 30 minutes 30 minutes Stabilization time typical 2 sec Weighing unit kg Auto Off Available options Ambient temperature 10 C 40 C Moist environment 0 95 non condensing Electric Supply Input volta...

Page 6: ...10 minutes 10 minutes Stabilization time typical 2 sec Weighing unit kg Auto Off Available options Ambient temperature 10 C 40 C Moist environment 0 95 non condensing Electric Supply Input voltage 11...

Page 7: ...30 minutes 10 minutes 10 minutes Stabilization time typical 2 sec Weighing unit kg Auto Off Available options Ambient temperature 10 C 40 C Moist environment 0 95 non condensing Electric Supply Input...

Page 8: ...inutes 10 minutes Stabilization time typical 2 sec Weighing unit kg Auto Off Available options Ambient temperature 10 C 40 C Moist environment 0 95 non condensing Electric Supply Input voltage 110 V 2...

Page 9: ...zation time typical 2 sec Weighing unit kg Auto Off Available options Ambient temperature 10 C 40 C Moist environment 0 95 non condensing Electric Supply Input voltage 110 V 230 V AC Power pack second...

Page 10: ...SFB BA e 1625 10 2 Appliance overview 1 Battery status display 2 Keyboard 3 Weight display 4 Tolerance tag see chap 7 7 5 Weighing unit 6 Levelling screw 7 Spirit level underneath weighing platform...

Page 11: ...forward in menu Display sum total Navigation key Digit selection to the right Add weighing value in summation memory Navigation key Digit selection to the left Calculate weighing data via interface C...

Page 12: ...rning to the first digit after the last digit has been pressed To change the selected flashing digit press repeatedly until the desired value appears Then select by using additional digits and change...

Page 13: ...his may damage the measuring system Impacts and overloading exceeding the stated maximum load max of the weighing plate minus a possibly existing tare load must be strictly avoided Balance may be dama...

Page 14: ...Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning even if you are already familiar with KERN balances 4 2 Personnel trainin...

Page 15: ...r humidity on the appliance may occur if a cold appliance is taken to a considerably warmer environment In this case acclimatize the disconnected appliance for ca 2 hours at room temperature Avoid sta...

Page 16: ...sealed screws must not be unscrewed Accurate weighing results require a weighing bridge with perfect horizontal alignment During initial installation and after each change of work area it is necessar...

Page 17: ...le example models platform size 300 x 240 mm Attach the tripod to the platform acc to fig using the 4 screws 1 securing disks and washers Ensure that the cable is not damaged nor squeezed Screw in sup...

Page 18: ...from holder for that remove the turning knobs 3 on the side Attach the tripod with the four raised counter sunk head screws 4 and the nuts on the holder of the display unit Re attach and position dis...

Page 19: ...SFB BA e 1625 19 Scope of delivery models platform size 400 x 300 mm Tripod tube Adapter display unit Tripod foot...

Page 20: ...he battery is getting low Approximately 10 h of instrument usage are left afterwards it will shut off automatically Use the supplied battery charger for charging the battery Charge status of rechargea...

Page 21: ...well as in case of fluctuating environment temperature To receive accurate measuring values it is also recommended to adjust the display unit periodically in weighing operation In weighing systems wit...

Page 22: ...must be re verified by an authorised agency and a new verification wire seal mark fitted before it can be reused for applications subject to verification Call up menu 1 Switch on balance and during th...

Page 23: ...s see chap 2 1 1 select the desired setting the active digit is flashing Acknowledge with Carefully place adjusting weight in the centre of the weighing plate Wait for stability display then press Aft...

Page 24: ...ing noLin by Ensure that there are no objects on the weighing plate Wait for stability display then press The currently set adjustment weight will be displayed To change by using the navigation button...

Page 25: ...ter testing instruments control The adjustment is locked for verified balances To disable the access lock destroy the seal and actuate the adjustment switch Position of the adjustment switch see chap...

Page 26: ...he centre of the weighing platform Wait for stability display then press When Ld 2 is displayed put the second adjustment weight 2 3 max carefully in the centre of the weighing platform Wait for stabi...

Page 27: ...a balance is used where obligation to verify exists as described above it must be verified and re verified at regular intervals Reverification is carried out according to the relevant national statuto...

Page 28: ...er the instrument can be reset to zero whenever needed as described below Remove load from weighing system Press and the zero display as well as the ZERO indicator will appear 7 4 Simple weighing Plac...

Page 29: ...off the displayed weighing unit press Acknowledge with The next unit with the current setting will be displayed To enable off disable on the displayed weighing unit press Acknowledge with Repeat seque...

Page 30: ...pacity is full Switch between gross weight and net weight by pressing the key To delete the tare value remove load from weighing plate and press 7 7 Weighing with tolerance range You may determine an...

Page 31: ...ance weighing may be set either by calling up menu block P0 CHK See chap 8 or by applying the faster option of pressing the key combination 7 7 1 Tolerance check for target weight Settings Press and a...

Page 32: ...akes place whether the goods to be weighed are within the two tolerance limits Weighing with tolerance range Tare when using a weighing container Put on goods to be weighed tolerance control is starte...

Page 33: ...ttons see chap 2 1 1 the currently enabled digit will be flashing Confirm input by Press repeatedly until is displayed Press the current setting for the upper limit will be displayed To enter the uppe...

Page 34: ...ng container Put on goods to be weighed tolerance control is started The signal lights indicate whether the load is within the two set limits Load below specified tolerance Load within specified toler...

Page 35: ...re weighed goods can only be added when the display zero Place goods to be weighed B Wait until the stability display appears then press The weight value will be added to the summation memory and poss...

Page 36: ...The data in the summation memory are deleted Printout example KERN YKB 01N verified weighing system Menu setting P1 COM or P2 COM Lab 2 Prt 7 Menu setting P1 COM or P2 COM Lab 0 Prt 0 1 First weighin...

Page 37: ...ed by printing Remove the weighed good More weighed goods can only be added when the display zero Place goods to be weighed B After the standstill control sounds a signal tone The weighing value is ad...

Page 38: ...eference quantity the higher the counting exactness In weighing mode press and hold until the message P 10 appears that is used to set the reference quantity Use to set the desired reference quantity...

Page 39: ...hap 8 or by using the faster option of a key combination The indicator shows HOLD as long as the animal weighing function remains enabled Place the load onto the weighing system and wait until it is f...

Page 40: ...key is pressed To cancel locking keep pressed and at the same time 2s until U LCK appears 7 13 Display background illumination Keep pressed 3s until setbl appears Press again and the current setting w...

Page 41: ...3s until setbl appears Call up AUTO OFF function using Press current setting appears Use to select desired setting of 0 AUTO OFF function disabled of 3 Weighing system will be turned off after 3 min...

Page 42: ...lp of the individual menu items can be selected one after the other Select setting Confirm selected menu item by pressing The current setting will be displayed Change settings To change to the availab...

Page 43: ...d 4d 0AUto Zero setting range Load range where the display after switching on the balance is set to zero Selectable 0 2 5 10 20 50 100 0rAGE Zero setting range Load range where the display is set to z...

Page 44: ...acity on Dual range balance R1 inc Readability 1st weighing range R1 cap Capacity 1st weighing range R2 inc Readability 2nd weighing range R2 cap Capacity 2nd weighing range CAL noLin For adjustment s...

Page 45: ...2 PCS L Lower limit value Tolerance check counting input see chap 7 7 2 BEEP no Acoustic signal for weighing with tolerance range switched off ok Audio sound when load is within tolerance limits ng A...

Page 46: ...P 1 Weighing scale capacity max 1 Weighing range CAP 2 Weighing scale capacity max 2 Weighing range CAL noLin Adjustment see chap 6 5 1 LinEr For linearization see chap 6 6 1 GrA Not documented dUAL 2...

Page 47: ...switched off P6 SP 7 5 15 30 Not documented Tab 1 Printout examples Standard printer Lab Prt 0 1 2 3 0 3 GS 5 000kg NT 5 000kg TW 5 000kg GW 10 000kg GS 5 000kg TOTAL 10 000kg NT 5 000kg TW 5 000kg GW...

Page 48: ...rsion Do not apply aggressive detergents solvents etc 9 2 Service maintenance The appliance may only be opened by trained service technicians who are authorized by KERN Before opening disconnect from...

Page 49: ...Keyboard error Err 6 Value outside the A D changer range Weighing plate not installed Damaged weighing cell Damaged electronics Err 9 Stability display does not appear Check the environmental conditio...

Page 50: ...hing balance to the interface of the printer Faultless operation requires an adequate KERN interface cable Communication parameters baud rate bits and parity of balance and printer must match For deta...

Page 51: ...BLE US UNSTABLE HEADER2 NT NET GS GROSS 10 5 Remote control instructions Command Function Printout examples S Stable weighing value for the weight is sent via the RS232 interface ST GS 1 000KG W Weigh...

Page 52: ...ted The displayed weight is permanently changing Draught air movement Table floor vibrations Weighing plate has contact with other objects Electromagnetic fields static charging choose different locat...

Page 53: ...ration of Conformity To view the current EC EU Declaration of Conformity go to www kern sohn com ce The scope of delivery for verified weighing balances conformity rated weighing balances includes a D...

Reviews: