6.7.2 Adjustment procedure (models PFB-M):
Observe stable environmental conditions.
A warming up time (see chapter 1; Technical Data) is required for stabilization.
Ensure that there are no objects on the weighing plate.
Balance display
Operation
Start balance by pressing
During the
selftest press
.
F1 Unt
Press
repeatedly
Press adjustment switch on the lower side of the balance!
Press
Enter password:
Press
,
,
subsequently.
Press
P2 CAL
Press
displayed.
Put the necessary adjustment weight (see
chap.1; Technical Data) on the centre of the
weighing plate.
Adjustment will be carried out automatically after dead
stop control.
Wait for stability display