PBJ_N-SH-e-1810.docx
24
9.3 Automatic adjustment by PSC (Perfect Self Calibration), only models PBJ
The PSC function determines the ambient temperature of the scale at any time. If the
tolerance groups are exceed upwards or downwards, this is signalled and the necessary
adjustment is carried out fully automatically. This will ensure that the scale is ready at any
time.
Activate function:
Press
3 times in weighing mode.
Menu group 1 „Adjustment“ is displayed, the indicator flashes.
Acknowledge by
, the current setting flashes.
Press
several times until „A“ flashes.
Acknowledge with
.
is used to change-over among the following settings.
„PSC on“
(menu element 5) = function activated
„PSC off“
(menu element 6) = function deactivated
The current setting is marked by the stability display (
).
Acknowledge selection by
.
Press
repeatedly or for 3 s and the scale returns into
weighing mode.
The flashing weight symbol
signals that automatic adjustment is imminent.
In order to avoid that adjustment is started possibly during a series of measurements,
actuate
as soon as the weight symbol flashes. This will interrupt automatic
adjustment.
If the PSC function has not been activated, the operator has to adjust by means of an
internal adjustment weight (chapter
) as soon as the weight symbol
flashes.