![KERN PBJ-N Service Manual Download Page 20](http://html1.mh-extra.com/html/kern/pbj-n/pbj-n_service-manual_1957049020.webp)
PBJ_N-SH-e-1810.docx
20
8
Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each
balance must be coordinated - in compliance with the underlying physical weighing principle -
to the existing acceleration due to gravity at its place of location (only if the balance has not
already been adjusted to the location in the factory). This adjustment process must be carried
out for the first commissioning, after each change of location as well as in case of fluctuating
environment temperature. To receive accurate measuring values it is also recommended to
adjust the balance periodically in weighing operation.
Observe stable environmental conditions. A warming up time (see
chapter 1) is required for stabilization. Ensure that there are no
objects on the weighing pan.
8.1 Manual adjustment by CAL button
The scales have been set by the manufacturer so that adjustment can be started directly from
the weighing mode using the
CAL
button.
Models PBJ: Adjustment with internal weight
Other adjustment procedures can be activated in the menu.
8.2 Adjustment with internal weight
Condition
:
Menu setting „
I.CAL
“ / Menu element 1
In weighing mode press
. I-CAL is displayed.
If „I-CAL“ is not displayed, press
to return into weighing mode and activate
menu element 1, see menu overview.
Summary of Contents for PBJ-N
Page 4: ...PBJ_N SH e 1810 docx 4 ...
Page 40: ...PBJ_N SH e 1810 docx 40 Fig 7 Fig 8 ...
Page 47: ...47 PBJ_N SH e 1810 docx Fig 22 Fig 23 ...
Page 55: ...55 PBJ_N SH e 1810 docx 23 Drawing Fig 31 ...
Page 56: ...PBJ_N SH e 1810 docx 56 ...
Page 57: ...57 PBJ_N SH e 1810 docx Fig 33 ...