Press the
[TARE]
key, the display will show the
present numerical value.
Then press the
[ENTER]
key.
Press the
[HOLD]
key.
Press the
[TARE]
key.
Enter the required calibration weight value (refer to
chapter 1, „Technical data”): Select the item to be
changed with the
HOLD
key and set its numerical
value with the
[TARE]
key.
Confirm by pressing the
[ENTER]
key.
Place the calibration weight carefully in the centre of
the scale plate, and the display will show a numerical
value. Press the
[ENTER]
key. The adjustment proc-
ess is started.
When the adjustment is finished successfully, the
scale is automatically switched over to the weighing
mode again and the calibration weight value will be
displayed.
Remove the calibration weight.
Note
:
In the case of verified scales, switch off a scale and set the adjustment switch in
verification position.