8 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each display unit with connected weighing plate must be coordinated - in compliance
with the underlying physical weighing principle - to the existing acceleration due to
gravity at its place of location (only if the weighing system has not already been
adjusted to the location in the factory). This adjustment process must be carried out
for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the display unit periodically in weighing operation.
Keep adjustment weight ready.
Incline the balance to a side.
Press the adjustment button at the lower side of the balance
and keep it pressed. During this time touch the weighing plate.
"CAL" appears on the display.
Place the balance again in a vertical position. Press the
weighing plate firmly until "0" appears flashing in the display.
Unload weighing plate. The required adjustment weight is
shown.
(example)
Place adjustment weight on the weighing plate. "CAL.F"
appears on the display.
Take away adjustment weight. Adjustment now has finished.
Should be displayed another message than "CAL.F", remove the adjustment weight