9 Error message
“Err“ on the display indicates that the weighing range has been exceeded. Please
take the weight off the scales.
10 Instant help
Help:
Fault
Possible cause
The displayed weight is
permanently changing
•
Draught/air movement
•
Table/floor vibrations
•
Weighing plate has contact with other objects
•
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is
obviously incorrect
•
The display of the balance is not at zero
•
Great fluctuations in temperature.
•
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
If the error message remains inform manufacturer.