18
KXS_KXG-TM-BA_IA-e-1932
7.7.2 Not verifiable weighing systems
For invoking menu item <P6ZCL> see chap. 11.1.
Ensure that there are no objects on the weighing pan before pressing
.
Wait for stability display, then press
.
The currently set adjustment weight will be displayed.
Either apply the displayed adjustment weight or make changes with the help of
the navigation keys (See chap.3.1). Confirm by
, „LoAd“ will be shown.
Carefully place adjustment weight in the centre of the weighing plate.
Wait for stability display, then press
.
After the adjustment the balance will carry out a self-test. Remove adjustment
weight
during
selftest, balance will return into weighing mode automatically.
Summary of Contents for KXG-TM
Page 6: ...5 KXS_KXG TM BA_IA d 1932 1 1 Abmessungen KXS TNM KXG TNM ...
Page 13: ...KXS_KXG TM BA_IA d 1932 12 3 3 Übersicht Anzeigezeichen ...
Page 61: ...KXS_KXG TM BA_IA d 1932 60 Kopplungscode des Gerätes eingeben anklicken Code 111111 eingeben ...
Page 62: ...61 KXS_KXG TM BA_IA d 1932 Weiter anklicken Schließen anklicken ...
Page 63: ...KXS_KXG TM BA_IA d 1932 62 Bluetooth Netzwerkgerät anzeigen Doppelklick ...
Page 75: ...KXS_KXG TM BA_IA e 1932 5 1 1 Dimensions KXS TNM KXG TNM ...
Page 81: ...KXS_KXG TM BA_IA e 1932 11 3 3 Overview display icons ...
Page 129: ...KXS_KXG TM BA_IA e 1932 59 Click on Enter pairing code of the device Enter code 111111 ...
Page 130: ...60 KXS_KXG TM BA_IA e 1932 Click on Next Click on Close ...
Page 131: ...KXS_KXG TM BA_IA e 1932 61 Display Bluetooth network appliance Double click ...
Page 143: ...KXS_KXG TM BA_IA f 1932 5 1 1 Dimensions KXS TNM KXG TNM ...
Page 149: ...KXS_KXG TM BA_IA f 1932 11 3 3 Aperçu caractères d affichage ...
Page 191: ...KXS_KXG TM BA_IA f 1932 53 12 RS 485 occupation pin ...
Page 200: ...62 KXS_KXG TM BA_IA f 1932 Cliquer sur poursuivre Cliquer sur clôturer ...
Page 201: ...KXS_KXG TM BA_IA f 1932 63 Afficher périphérique réseau du Bluetooth Double clic ...