Guidance and manufacturer’s declaration-electromagnetic immunity
The MBB15K2DM is intended for use in the electromagnetic environment specified below.
The customer or the user of the MBB15K2DM; should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment-guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast
transient/burst IEC
61000-4-4
± 2kV for power
supply lines + 1kV
for input/output
lines
± 2kV for power
supply lines Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-5
± 1kV line(s) to
line(s) ± 2kV
line(s) to earth
± 1kV differential
mode Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT(>95% dip
in UT) for 0,5
cycle 40%
UT(60% dip in UT)
for 5 cycles 70%
UT(30% dip in UT)
for 25 cycles <5%
UT(>95% dip in
UT) for 5 s
<5% UT(>95% dip
in UT) for 0,5
cycle 40%
UT(60% dip in UT)
for 5 cycles 70%
UT(30% dip in UT)
for 25 cycles <5%
UT(>95% dip in
UT) for 5 s
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the MBB15K2DM requires
continued operation during
power mains interruptions,
it is recommended that the
MBB15K2DM be powered
from an uninterruptible
power supply or a battery.
Power
frequency(50/60 Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
The MBB15K2DM power
frequency magnetic fields
should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Summary of Contents for KERN MBB-M series
Page 41: ...MBB M BA d 1617 40 Die Auswahl erfolgt mit der HOLD und TARE Taste Werkseinstellung ...
Page 82: ...40 MBB M BA e 1617 Press the HOLD and TARE buttons to make the selection default setting ...
Page 85: ......
Page 107: ...MBB M BA f 1617 23 Branchement CN 3 de pile rechargeable Branchement CN 4 de piles AA x 6 ...
Page 125: ...MBB M BA f 1617 41 La sélection s opère sur la touche HOLD et TARE réglé en usine ...