12
- GAB/GAS-SH-e-0711
3) Operation
Operation
Display
1. Stating
Test
X2
X1
2. Scale No. and Decimal point indication mode
- Press the [ TARE ] key
→
1
st
digit flashes
- Example: ASIA 6kg (5g/2g) Multi interval X1=2, X2=4
2.1 Scale No. Setting
- Press the [ MODE ] key one time (X1=2)
X2
X1
2.2 Decimal point indication Setting
- Press the [ ZERO ] key
→
2
nd
digit flashes
X2
X1
- Press the [ MODE ] key three times (X2=4)
X2
X1
- When the C3 mode is required, press the TARE key.
- When finishing, press the Memory Switch to
record the data to the memory, and enter the Ending
Test Mode.
2.1.8 C3 Mode – Span Adjustment
1) Operation
Operation
Display
1. Stating
Test
2. Span Adjustment mode
2.1 Press the TARE key two times.
2.2 Press the MODE key
→
displays original A/D data
(The normal range for original A/D data is 1000 to
25000 counts).
2.3 Press the MODE key without load on the weigh platter
→
Approx. 5000 count is diplayed.