13
7. Display LIN
Press the
key
8. Display CAL (adjustment by means of adjusting weights)
Press the
key. SET PL will appear on the display screen.
Press the
key in ordfer to start the calibration procedure.
You will subsequently be asked to adjust your weighing system at a certain value
In the event that you do
not
wish to make the adjustment using this value
(e.g. because you do not have sufficient weights at your disposal, keep pressing the
key
until the desired adjusting weight appears on the display screen (the adjusting weight is
reduced each time the key is pressed).
Note
: It is recommended in general to adjust the balance when
fully loaded
Apply the selected adjusting weight and use the
key to confirm.
Successful completion of adjustment is confirmed when done appears on the display screen.
END
will appear automatically on the display screen (now see item 9).
9. Display END
Remove the adjusting weights from the weighing plate!
Press the
key order to save the alterations.
Store
will appear on the display screen.
Press the
key to confirm.
Following this the balance will return to weighing mode.
After you exit from the service mode, the scale must be switched off and then
back on again!