![KERN FOB 500-1S Service Manual Download Page 6](http://html1.mh-extra.com/html/kern/fob-500-1s/fob-500-1s_service-manual_1957252006.webp)
6
FOB-S-e-1310
5 Spare Part List
Beschreibung
Description
No. In Drawing
KERN Nummer/KERN
part no.
Stellfuss
Footstand
11 RFOB-S-8002
Wägeplatte
Scale pan
1 RFOB-S-6001
Folientastatur
Key panel of FOB500K-1S
6
RFOB-S-4001
Folientastatur
Key panel of FOB5K-1S
RFOB-S-4002
Anschlußbuchse
Steckernetzteil
Power socket
8 RFOB-S-5001
6 ERROR MESSAGE
Error
message
Description
Solution
Nothing appears on the
display after pushing ON/OFF
key.
1. Check display.
2. Disassemble indicator.
3. Check wires and control
panel.
Can’t switch on scale using
AC adaptor.
1. AC adaptor damaged.
→Replace adaptor.
2. AC jack wires are
disconnected.
Low Battery
This warning shows that the
voltage of battery is too low to
use.
Please replace new batteries
or plug the AC adaptor for
operation.
Counting error (too low)
The loading is lower than limit
when power on.
1. Please increase the
loading and try again.
2. Check the load cell and its
wiring.
3. Replace new deck.
Counting error (too high)
Scales zeroed exceed its
initial balance.
1. Remove the weight from
scale.
2.
Check the load cell’s
resistance.
3. Replace new deck.
Overload
The loading is over limit when
power on, please reduce the
loading and try again.
Remove the weight from
scale.
Overload: Max. Ca 9d.
EEPROM error
The program of the scale has
error.
Replace controller PCB, and
re-calibrate the scale.