![KERN FOB 0.5K-4NS Operating Manual Download Page 31](http://html1.mh-extra.com/html/kern/fob-0-5k-4ns/fob-0-5k-4ns_operating-manual_1957249031.webp)
FOB-BA-e-1820
31
9.4 Hold function (FOB, FOB-NL/-LM/ models only))
The balance has an integrated standstill function (mean value calculation).
With this it is possible to weigh unstable samples to be weighed exactly.
There are 2 possibilities:
Balance display
Operation
•
Turn on balance by pressing the
ON/OFF
key.
The balance will carry out a self-test.
•
Wait for „
0
“ display
(example)
Put the weighed item in place and press the
HOLD
key; "HOLD" will appear and a small triangle
flashes. During this time, an average value is
recorded.
The triangle will stop flashing, the stability display
appears and the weight value obtained is displayed.
By pressing the
HOLD
button several times, the
balance returns to weighing mode. The triangle
disappears.
9.5 Display background illumination (FOB, FOB-NL/-LM/ models only)
The balance has the ability to turn the backlight of the display on or off.
Press the
TARE
key for approx. 3 sec.