EOS_F-BA-e-1310
15
In weighing mode press the
UNIT
key and keep it pressed until „
CAL
“ followed
by the weight value of the required adjustment weight in displayed. „
LOAd
“ is
displayed alternatively.
(Example)
Carefully place the required adjustment weight (see tab. 1 „Technical data“) in
the centre of the weighing plate
Press
TARE
key, adjustment is started.
(Example)
Wait until the stability display appears, after that the adjustment is finished. The
balance will automatically return to Weighing mode upon successful completion
of the adjustment and will display the weight of the adjustment weight.
(Example)
Take away adjustment weight.
An error message will be displayed in the event of an adjustment error or
incorrect adjustment weight. Remove the adjustment weight and repeat the
adjustment process.
Keep the adjustment close to the balance. Daily control of the weighing
exactness is recommended for quality-relevant applications.