KERN
EMS 6K0.1
EMS 6K1
EMS 12K0.1
EMS 12K1
Readability (d)
0.1 g
1 g
0.1 g
1 g
Weighing range (max)
6 kg
6 kg
12 kg
12 kg
Taring range (subtractive)
6 kg
6 kg
12 kg
12 kg
Reproducibility
0.1 g
1 g
0.1 g
1 g
Linearity
± 0.3 g
± 3 g
± 0.3 g
± 3 g
Minimum unit weight at
piece counting
0.2 g
2 g
0.2 g
2 g
Warm-up time
120 min
30 min
120 min
30 min
Reference quantities at
piece counting
5, 10, 20, 25, 50
Weighing Units
dwt, g, oz, ozt
Recommended adjustment
weight, not added (class)
6 kg (F2)
6 kg (M1)
12 kg (F2)
12 kg (M1)
Stabilization time (typical)
3 sec.
Operating temperature
+ 5° C .... + 35° C
Humidity of air
max. 80 % (not condensing)
Housing (B x D x H) mm
200 x 280 x 63
Weighing plate mm
160 x 160
Weight kg (net)
1,4
Input Voltage
110V-230V AC
Power pack
secondary voltage
9 V, 300mA
Battery operation
9 V compound battery (optional)
Operating period: 40 h
Auto Off
3 min.
EMS-BA-e-1514
5