background image

 

4

 

 

EHA-BA-e-1910

 

 

5.  Operation 

Weighing: 

 

 

Place the balance on a firm, level surface. Level balance with foot 
screws until the air bubble of the water balance is in the prescribed 
circle. Check levelling regularly

 

 

 

To switch-on, press the 

ON/OFF

 button, the maximum weight of the balance will be displayed for a 

short time. After that the weight display appears, the balance is now ready for weighing.  
If required, set to zero by using 

TARE

 button. 

 

Place load on pan and read the weighing result. 

 

Use the 

MODE

 button to switch over into the available weighing settings. 

 

                                

 

 

Taring: 

 

Place the empty weighing container. 

 

Press the 

TARE

 button, the zero display disappears. 

 

Weigh the material, the net weight will be indicated. 

 

When the balance is unloaded the saved taring value is displayed with negative sign. 

 

To delete the stored tare value, remove load from weighing pan and press the 

TARE

 button.  

 

Parts counting: 

 

In weighing mode press the 

PCS

 button for 2 seconds, the balance changes into counting mode, 

25 PCS will be displayed. 

 

Press the 

MODE

 button to select the desired reference number of pieces (selectable options 

include 25, 50, 75, 100) 

 

While PCS is flashing, place a reference which corresponds to the selected number of pieces 

 

Press PCS again 

 

Remove reference. 

 

If necessary, tare the empty balance container. 

 

Place counted pieces on pan and read the quantity of pieces. 

 

Use 

MODE

 button to return to the weighing mode.  

 

 

6.  Important Notes 

 

Do not overload balance.

 

 

The occurrence of electromagnetic fields may result in major display deviations. Then change 
location.

 

 

Avoid high air humidity, vapours and dust. 

 

 

Do not bring the balance directly in contact with liquids. 

 

 

Do not leave permanent load on the weighing pan. 

 

 

Never operate balance in explosive environment. The serial version is not  
explosion protected.

 

 

Check the balance regularly with external test weights.

 

 

Warranty claims will be voided when the housing is opened. 

 
 

 