EW/EG-N/EWB-e-1627
23
Turn the adjusting knob on the right side of the
balance to the
WEIGH
position.
Calibration is now finalised.
The balance will automatically return to weigh-
ing mode.
5.8.3 Adjusting with external weight (only KERN EW and EWB)
Carry out adjustment with the recommended adjusting weight (see Chapter 1 „Tech-
nical Data“). The adjustment can also be carried out with different adjusting weights
(see table), but not ideal from a metrological point of view. Information concerning
the adjusting weights is available at:
http://www.kern-sohn.com
Model
Alternative adjusting weight
EW 220-3NM
100 g
EW 420-3NM
100 g
EW 620-3NM
200 g
EW 820-2NM
200 g
EW 2200-2NM
500 g
EW 4200-2NM
1000 g
EW 6200-2NM
2000 g
EW 12000-1NM
5000 g
EWB 220-2M
100 g
EWB 620-2M
200 g
EWB 1200-1M
500 g
CAL
buSY
End.