7.4  Dosage und Zero-tracking 

When the Auto-Zero-function is activated, any slight changes of the zero readout are 
automatically tared. 
In the event that small quantities are removed or added to the material to be 
weighed, incorrect weighing results can be displayed due to the “stability compensa-
tion“ in the balance. (Example: Slowly draining fluids from a container on the bal-
ance.)   
Having dosages with slight changes of weight it is recommended to deactivate this 
function. 

 

With switched off

 Zero-Tracking

 the display however is more unstably. 

 

Activate/deactivate Zero-Tracking

 

Ö

  When the balance is switched off, press the 

TARE

 -key and keep it pressed 

Ö

  Do not release the 

TARE

 -key. Press also the 

ON/OFF

 -key and keep it pressed  

Ö

 Keep the 

ON/OFF

 -key pressed, however release the 

TARE

 -key 

Ö

 Press 

TARE

 -key anew 

Ö

  Keep both keys pressed until in the display “

bl

” appears 

Ö

  Release both keys. The balance is situated in the menu. 

Ö

 Press the 

ON/OFF

 -key until „

tr

“ appears  

Ö

  Confirm with the 

TARE

 key  

Ö

 Press the 

ON/OFF

 key to switch among the settings: 

Display 

Mode 

„tr“ ON

 

Auto-Zero on 

„tr“ OFF

 

Auto-Zero off 

 

Ö

  Confirm with the 

TARE

 key. Wait until the balance automatically returns to weigh-

ing mode 

 

Summary of Contents for ECB

Page 1: ...mbH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating instruction Platform scale KERN ECB Version 1 3 05 2007 GB EC...

Page 2: ...installation and commissioning _________________________ 8 6 1 Place of installation place of use ____________________________________ 8 6 2 Unpacking the Balance _____________________________________...

Page 3: ...________________________________ 17 8 1 Cleaning _______________________________________________________ 17 8 2 Maintenance upkeep _____________________________________________ 17 8 3 Disposal________...

Page 4: ...ing range subtractive 10 kg 20 kg 50 kg Reproducibility 10 g 20 g 50 g Linearity 20 g 40 g 100 g Adjustment tolerance as per M3 5 kg 10 kg 20 kg Stabilisation time typical 2 3 sec Optimal temp of oper...

Page 5: ...n Normen bereinstimmt Fran ais Nous d clarons avec cela responsabilit que le produit auquel se rapporte la pr sente d claration est conforme aux normes cit es ci apr s Espa ol Manifestamos en la prese...

Page 6: ...ing equipment Be sure to avoid impact shock and overloading the balance in excess of the pre scribed maximum load rating max minus any possible tare weight that is already present This could cause da...

Page 7: ...mal 4 Fundamental safety information 4 1 Observe the information in the operating instructions Please read the operating instructions carefully before erecting and commissioning even if you already ha...

Page 8: ...rs and dust Do not expose the device to extreme dampness for longer periods of time In admissible bedewing condensation of air moisture on the device can occur if a cold device is taken into a signifi...

Page 9: ...Insert 6 x 1 5volt AA batte ries Replace the battery cover To preserve battery life the balance switches off automatically after 3 minutes if no weighing has taken place For battery operation the bala...

Page 10: ...ECB BA e 0713 10 Rechargeable Battery Installation 1 Remove battery cover 2 Slide spacer to left or right lever down on corner then remove the spacer 3 Insert and connect rechargeable battery...

Page 11: ...alance periodically during weighing operation in order to obtain exact measured values 6 6 Adjusting Using a precision weight the accuracy of the balance can be checked at any time and adjusted Adjust...

Page 12: ...heavier than the weighing range the display indicates E Overload The balance is switched off by pressing the ON OFF key once 7 2 Taring Switch the balance with the ON OFF key on and wait until displa...

Page 13: ...ew Keep both keys pressed until in the display bl appears Release both keys The balance is situated in the menu Press the ON OFF key until bl appears Confirm with the TARE key Press the ON OFF key to...

Page 14: ...h switched off Zero Tracking the display however is more unstably Activate deactivate Zero Tracking When the balance is switched off press the TARE key and keep it pressed Do not release the TARE key...

Page 15: ...it pressed Keep the ON OFF key pressed however release the TARE key Press TARE key anew Keep both keys pressed until in the display bl appears Release both keys The balance is situated in the menu Pre...

Page 16: ...wever release the TARE key Press TARE key anew Keep both keys pressed until in the display bl appears Release both keys The balance is situated in the menu Press the ON OFF key until rSt appears Confi...

Page 17: ...and rub off using a clean soft cloth Loose sample residue powder can be removed carefully using a brush or hand vac uum cleaner Remove any spilt material to be weighed immediately 8 2 Maintenance upk...

Page 18: ...The weight display changes continu ally Draught air movement Table floor vibrations The weighing plate is in contact with for eign matter Electromagnetic fields static charging choose different locati...

Reviews: