ALJ-A03-BA-e-1810 

 

 

Always use the OEM power adapter. The printed volt-
age value must comply with the local voltage. 

 

Hazard of harm, ion source blades are very sharp. 

 

The ionizer generates poisonous ozone, ensure suitable 
ventilation. 

 

Before you start any maintenance or cleaning, discon-
nect the ionizer from the power supply. 

 

Always disconnect the unused ionizer from the power 
supply. 

 

CAUTION

 

 

Always maintain and clean the ionizer at regular inter-
vals. 

●  Ion source cleaning following 100 hours. 
●  Ion source replacement following 10,000 hours. 

 

Operating the damaged ionizer can result in short cir-
cuit, fire or electric shock.  

 

It is prohibited to switch it outdoors or in vehicles, this 
shall invalidate any warranty. 

 

If any electromagnetic fields are present, high readout 
fluctuations are possible (mistaken weighing results). 
Discharge the sample at the appropriate distance from 
the balance. 

 

The ionizer may heat slightly during the operation. 

 

For improved ionization results, switch the blower on. 
This will make the sample discharge time shorter. 

 
 
 
 

Summary of Contents for ALJ-A03

Page 1: ...Sohn GmbH Ziegelei 1 D 72336 Balingen E mail info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 www www kern sohn com User Manual Ionizer KERN ALJ A03 Version 1 0 2018 08 GB ALJ A03 BA e...

Page 2: ...RMATION 3 2 BASIC SAFETY GUIDELINES 3 3 TECHNICAL SPECIFICATION 5 4 DEVICE OVERVIEW 6 5 UNPACKING PLACEMENT AND START UP 7 Unpacking 7 Scope of delivery 7 Installation 7 Power 8 6 START UP 8 Applicati...

Page 3: ...ith explosion haz ard present Serial installation is not an explosion proof one Protect the ionizer from high ambient humidi ty temperature vapours and dust Ensure location free from water oil Do not...

Page 4: ...e ionizer at regular inter vals Ion source cleaning following 100 hours Ion source replacement following 10 000 hours Operating the damaged ionizer can result in short cir cuit fire or electric shock...

Page 5: ...n source Weight 525 g Dimensions cm 110 105 60 Ambient conditions 0 50 C air humidity 20 80 non condensing Power adapter Input voltage 100 240 VAC 50 60 Hz Ionizer input voltage 12 VDC 500 mA Pollutio...

Page 6: ...automatic switch off in 8 h 2 Active ion source Time Mode t2 minute work mode To switch between the two work modes press one of the two buttons Indication overview Green LED Ionizer operating Ionizer...

Page 7: ...lacement and start up Unpacking Remove the device carefully from its packaging remove the plastic bag and place the device in the working loca tion Scope of delivery 1 Ionizer 2 Power adapter EURO UK...

Page 8: ...connect the ionizer to the power adapter when the device is off 6 Start up Switch the ionizer on pressing button The ionizer is in the Continuous Mode the working time is 8 hours Red LED is on The ion...

Page 9: ...ischarge time shorter Discharging powder samples Discharging prevents whirls a problem for toxic samples If the samples easily whirl switch the blower off Discharging weighed material glass windproof...

Page 10: ...se a soft cloth to wipe the device dry after cleaning Keep the ion sources clean Maintenance service The device can be operated and maintained solely by ser vice technicians trained and authorised by...

Reviews: