background image

OBN-13-BA-e-1510 

 

  12 

4.5  Condenser 

 
We  recommend  that  you  use  the  course  adjustment  knob  to  bring  the  specimen 
stage  to  its  uppermost  position.  Use  the  focus  dial  of  the  condenser  to  move  the 
condenser  holder  to  the  central  position.  In  this  way  the  condenser  can  be  fitted  at 
the  right  place  in  the  condenser  holder  and  fixed  with  the  adjusting  screw.  When 
doing  this,  you  should  be  able  to  read  the  scale  from  the  front.  You  should  avoid 
touching the optical lenses with bare fingers. 
 
 

For points 6 (Polarisation unit) and 7 (Camera connection), please see Chapter 
8 Use of optional accessories. 

 
 
 
 
 

Operation 

 

5.1  Getting started 

 
The  very  first  step  is  to  establish  a  power  connection  using  the  mains  plug.  You 
should  first  adjust  the 

dimmer

  to  a 

low  level

,  so  that  when  you  look  through  the 

eyepiece  for  the  first  time,  your  eyes  are  not  immediately  subject  to  a  high  level  of 
light. You can now switch on the 

lighting

 using the 

main switch

 

The next step is to 

place an object holder

 with the sample on the angle table. To do 

this,  the  cover  glass  must  be  facing  upwards.  You  can  fix  the  object  holder  to  the 
table using the slide holder 

(see illustration on the left)

. To move the sample into the 

beam path, you must use the adjustment wheels on the  right of the angle table 

(see 

illustration on the right)

.  

You can position a total of two object holders at the same time. 
 

 

 

Summary of Contents for OBN-13

Page 1: ...Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Transmitted light laboratory microscope KERN OBN 13 OBN...

Page 2: ......

Page 3: ...Condenser 12 5 Operation 12 5 1 Getting started 12 5 2 Pre focussing 13 5 3 Adjusting the interpupillary distance 14 5 4 Dioptre adjustment 14 5 5 Adjusting the magnification 15 5 6 Adjusting the Koeh...

Page 4: ...n additional risk of triggering an electric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The infor...

Page 5: ...ich are too high or too low vibrations dust or a high level of humidity The ideal temperature range is between 0 and 40 C and a relative humidity of 85 should not be exceeded The device should always...

Page 6: ...s these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organ...

Page 7: ...erture adjustment Condenser Condenser Field lens with Field diaphragm Control dial X Y axis Specimen stage Fine adjustment knob Coarse adjustment knob Dimmer Tube connectors with dioptre adjustment ri...

Page 8: ...7 OBN 13 BA e 1510 Rear view Adjusting ring torque Carrying handle Main switch Power connection Coarse adjustme nt knob Fine adjustment knob Fuse Condenser Focus dial...

Page 9: ...nity Trinokular 6V 20W Halogen Transmitting OBN 135 Infinity Trinokular 3W LED Transmitting Eyepieces WF 10x 20 mm Objectives 4x 10x 20x 40x 100x Product dimensions 390x200x400 mm Packaging dimensions...

Page 10: ...9 OBN 13 BA e 1510...

Page 11: ...OBN 13 BA e 1510 10 4 Assembly...

Page 12: ...ongest magnification appears You must make sure that you do not touch the lenses with your bare fingers and that no dust enters the apertures For objectives which are marked OIL you must use an immers...

Page 13: ...ccessories 5 Operation 5 1 Getting started The very first step is to establish a power connection using the mains plug You should first adjust the dimmer to a low level so that when you look through t...

Page 14: ...soon as an image is recognisable no matter how sharp then you should only adjust the focus using the fine adjustment knob Adjusting the torque of the coarse and fine adjustment knob Next to the left a...

Page 15: ...llary distance is set correctly 5 4 Dioptre adjustment The eye strengths of each eye of the microscope user can often be slightly different which in daily life has no consequences But when using a mic...

Page 16: ...objectives you should use the silver ring above the objectives see illustration When rotating the nosepiece you must always make sure that the objective which is about to be positioned in the beam pat...

Page 17: ...hragm When adjusting the Koehler lighting for the first time you must first select the lowest possible objective magnification so that you can carry out the following steps 1 Use the condenser focus w...

Page 18: ...he condenser up so high that it collides with the object holder Therefore care is needed when doing this 4 Use the centring screws of the condenser holder to bring the image from the field diaphragm i...

Page 19: ...3 of the pupil diameter If the eyepiece should be removed for checking then please make sure that no dirt or dust falls into the tube 7 It is possible to alter the brightness of the bulb using the di...

Page 20: ...sses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We would theref...

Page 21: ...Lower the specimen stage and position the 100x objective in the beam path 3 Bring the specimen stage or object slide to the objective very slowly until there is slight contact 4 Observe the object Th...

Page 22: ...lb can be pulled out of the socket and be replaced with a new one After the bulb holder has been replaced in the underside of the device and the screws replaced the bulb replacement procedure is compl...

Page 23: ...ece see illustration on page 10 where you can simply insert the analyser see left illustration The polariser on the other hand is only placed on the ring holder of the field lens see right illustratio...

Page 24: ...the microscope head must be pulled out The light from the microscope lighting is deflected so that it is completely in the beam path for the camera which causes a dark field of view in the eyepieces T...

Page 25: ...This consists of a simple PH condenser a PH objective with a specific magnification 10x 20x 40x or 100x a PH slider which is adapted to the lens being used a centring telescope and two green filters...

Page 26: ...es of the user using a green filter can produce a more effective and pleasant image To achieve this the green filter must be screwed onto the underside of the PH condenser 2 Comprehensive phase contra...

Page 27: ...a black objective ring The black one is central and the white one possible is pushed to one side see left illustration g Now press the turnscrews fitted on the side of the condenser towards the centr...

Page 28: ...tained in the spectrum of the light source HBO or LED The excitation filter is only permeable for the according wave length After that the excitation light hits a dichroic mirror which reflects it tow...

Page 29: ...OBN 13 BA e 1510 28 Overview of wave lengths for excitation and emission per excitation filter Blue line Wave length excitation Green line Wave length emission...

Page 30: ...29 OBN 13 BA e 1510...

Page 31: ...OBN 13 BA e 1510 30...

Page 32: ...er for FL module Fixing Allen screw for microscope head Main body Connection point for microscope head Fixing screw for lamp housing 7 8 9 10 11 HBO LED lamp housing Power connection Screw for attachi...

Page 33: ...device for field diaphragm Control lever for field diaphragm Centring device for aperture diaphragm Control lever for aperture diaphragm 6 7 8 9 10 Condenser control Screw for cover of lamp housing Ce...

Page 34: ...lected light units available for the microscopes of the series OBN 13 They have the following specifications Article number Illumination Excitation filter Centring objective Input voltage power supply...

Page 35: ...he reflected light unit for special fluorescence applications it will be favourable to adjust the microscope in the bright field mode This includes the placing of the sample adjusting the interpupilla...

Page 36: ...reduction of the lamps life time Equally the lamp must not be switched on directly after a switch off process When pausing the observation you should always slide in the control lever for the illumin...

Page 37: ...Use the two centring screws on the lamp housing for moving the light spot into the centre of the hair cross Warning Heat development on the lamp housing Changing the lamp HBO 1 Disconnect the FL refl...

Page 38: ...he mains power supply socket With the device switched off and unplugged you can screw out the housing The defective fuse can be removed from its housing and be replaced with a new one After that you j...

Page 39: ...ly centred The condenser is too low The field of view is dark or is not correctly illuminated The objective is not positioned correctly on the beam path The selector switch for the beam path is betwee...

Page 40: ...Dirt dust on the front lens of the condenser Dirt dust on the object One side of the image is blurred The stage was not correctly fitted The objective is not positioned correctly on the beam path The...

Page 41: ...y materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Reviews: