KLN 750W 121912
3-9
3.3.8
OUTPUT MODE (VOLTAGE SET: DIRECT OR ENTER)
There are two ways to program output voltage: DIRECT or ENTER. When output mode is set to
DIRECT, rotating the encoder with the output set to ON changes the output immediately as the
encoder is turned. When set to ENTER, the settings are not applied to the output until confirmed
by pressing ENTER.
1. Enter the menu (see PAR. 3.2.2 and Table 3-3) and rotate the encoder until Voltage display
shows
(OUT) to access OUTPUT mode.
2. Press ENTER key to view setting.
DIRECT mode displayed as
(DIR) in Current display.
ENTER mode displayed as
(ENT) in Current display.
3. Rotate encoder until Current display shows desired setting, then press ENTER to confirm.
3.3.9
POWER ON SETTING
When the unit is powered on or recovering from a loss of mains supply, the output can be con-
figured to restart with output status LAST or OFF. When set to LAST, output status is restored to
the last output status (ON or OFF) before the unit was turned off. When set to OFF, the output
status will always be off upon power supply restart.
1. Enter the menu (see PAR. 3.2.2 and Table 3-3) and rotate the encoder until Voltage display
shows
(P.ON) to access POWER ON setting.
2. Press ENTER key to view setting.
LAST mode displayed as
(LAST) in Current display.
OFF mode displayed as
OFF) in Current display.
3. Rotate encoder until Current display shows desired setting, then press ENTER to confirm.
3.3.10
RISE TIME (RAMP UP)
When the output is enabled, the time for the output to go from 0 to the programmed voltage is
referred to as the ramp up (rise) time, and is adjustable from 00.0 to 99.9 seconds. The time set-
ting affects the slope of the increase in output voltage.
1. Enter the menu (see PAR. 3.2.2 and Table 3-3) and rotate the encoder until Voltage display
shows
(R.UPT) to access RAMP UP setting.
2. Press ENTER key to view setting.
Ramp time displayed as
(0.00) in Current display.
3. Rotate encoder until Current display shows desired setting, then press ENTER to confirm.