BIT 4886 051010
B-21
B.80 SYSTem:ERRor? QUERY
SYST:ERR?
Syntax:
Short Form: SYST:ERR?
Long Form: SYSTem:ERRor?
Return Value:
<int_value,string>
Description:
Posts error messages to the output queue.
Returns the next error number followed by its corre-
sponding error message string from the instrument error queue. The error queue is a FIFO (first in,
first out) buffer that stores errors as they occur. As it is read, each error is removed from the queue
and the next error message is made available. When all errors have been read, the query returns
0,"No error". If more errors are accumulated than the queue can hold, it will overflow. The oldest errors
stay in the queue but the most recent errors are discarded. The last error in the queue will be
-350,"Too many errors." Error messages are defined in Table B-6.
B.81 SYSTem:ERRor:CODE? QUERY
SYST:ERR:CODE?
Syntax
:
Short Form: SYST:ERR:CODE?]
Long Form: SYSTem:ERRor:CODE?
Description:
Returns the three character error code without the ASCII definition string.
The error codes are
defined in table B-6 (See example, Figure B-8.)
B.82 SYSTem:ERRor:CODE:ALL? QUERY
SYST:ERR:CODE:ALL?
Syntax
:
Short Form: SYST:ERR:CODE:ALL?]
Long Form: SYSTem:ERRor:CODE:ALL?
Return Value:
Description:
Returns a comma-separated list of all error codes.
A maximum of 15 codes will be returned; if the
queue is empty, the power supply returns 0.
B.83 SYSTem:PASSword:CENable COMMAND
SYST:PASS:CEN
Syntax
:
Short Form: SYST:PASS:CEN <val>
Long Form: SYSTem:PASSword:CENable <val>
Description:
Sets the password enable state if the value matches the current password.
This command
allows other commands such as DIAG:SAV and CALibrate to operate.
B.84 SYSTem:PASSword:CDISable COMMAND
SYST:PASS:CDIS
Syntax
:
Short Form: SYST:PASS:CDIS <val>
Long Form: SYSTem:PASSword:CDISable <val>
Description:
Clears the password enable state if the <value> matches the current password.
B.85 SYSTem:PASSword:NEW COMMAND
SYST:PASS:NEW
Syntax
:
Short Form: SYST:PASS:NEW <old password>,<new password>
Long Form: SYSTem:PASSword:NEW <old password>,<new password>
Description:
Establishes new password.
The old (current) password is checked, then replaced by the new pass-
word. (See example, Figure B-8.)
B.86 SYSTem:PASSword:STATe? QUERY
SYST:PASS:STAT?
Syntax
:
Short Form: SYST:PASS:STAT?
Long Form: SYSTem:PASSword:STATe?
Return Value: <int_value> 0 or 1
Description:
Returns a 1 if the password state is enabled or a 0 if it is disabled.
B.87 SYSTem:REMote
COMMAND
SYST:REM
Syntax
:
Short Form: SYST:REM {ON | OFF} or {1 | 0}
Long Form: SYSTem:REMote {ON | OFF} or {1 | 0}
Description:
Used during serial (RS 232) communication to set the unit to remote (1 or ON) or local (0 or
OFF) mode.
This command must be issued prior to commands that affect the power supply output
(e.g., VOLT 10;:OUTP ON) to ensure the unit is in remote mode. See PAR. 4.5.3 and Figure B-7.
B.88 SYSTem:REMote? QUERY
SYST:REM?
Syntax
:
Short Form: SYST:REM?
Long Form: SYSTem:REMote?
Return Value: {1 | 0}
Description:
Identifies whether unit it in remote mode (1) or local mode (0) during serial (RS 232) communi-
cation.