1V range
VR101
RANGE: 1V
1) Input 1kHz (or 400 Hz), 1 Vrms sine wave, and set the pointer to 10.0 V of the 0-
10 scale. Check that the variable range is less than 98% and more than 102%
with respect to 10.0 (full-scale).
2) Waveforms shown on the oscilloscope shall not be deformed.
VT-181/VT-181E
8
ADJUSTMENT
ITEM
100 kHz
TC101
frequency
RANGE: 1V
characteristics
1) Input a 1kHz (or 400 Hz), 1 Vrms sine wave, and adjust the oscillator output so
that the pointer of the set points at 9.0.
2) Adjust the TC so that the pointer points at 9.0 when the frequency is changed to
100 kHz while the oscillator output remains unchanged.
PROCEDURE
ADJUSTMENT POINT
0
3
10
Scale 9
Scale 0-10
Scale 0-3
VT-181/VT-181E(K)
INT
98.4.9
10:31
AM
y [ W
15