background image

6

Kentech Instruments Ltd., Unit 9, Hall Farm Workshops, South Moreton, Didcot, Oxon, OX11 9AG, U.K.

10th. July 1999

3

Orientate the cell for maximum extinction. Ideally this should be done by looking at the isogyre
pattern obtained by scattering the incoming light over a range of angles. This can be done with
suitable tape over the input face of the cell. The laser should pass through the centre of the pattern.
Don’t forget to remove the tape!

4

Repeat 3 with the second cell.

5

Rotate the output polariser so that it is parallel to the input polariser.

6

Apply the relevant half wave voltage DC to on cell and repeat 3. The position should be near to the
existing position. Adjust the voltage so that the extinction is best. Note that misalignment and a
higher voltage can achieve an extinction minimum also and one is looking for the minimum
transmission near the existing position and at the lowest voltage.

7

Remove the voltage from the cell and repeat 3 on the second cell.

8

Remove the voltage from the second cell and now connect the quarter wave voltage to both cells.
Now work with small movements of the cells to obtain the best extinction ratio. The ratio is easily
obtained from the transmission obtained when the voltage is removed to the transmission when the
voltage is applied. Note that some power supplies will not discharge the cells when they are turned
off. It may be necessary to discharge the cell capacitance with a resistor (note that the capacitance
of any cabling to the cell will be much bigger than the cell and will store a significant amount of
energy.)

9

Remove the DC supply and connect the pulser. The light source may now be pulsed or DC to look
for the relevant timing of the pulsers, the laser source and the detector display.

Figure 2

A plot of DC transmission against applied voltage. The voltage is
applied to both cells simultaneously.

5.2

TIMING ADJUSTMENT

The relative timing of the four channels of the pulser must be set to obtain the best overall risetime.

It is assumed that the above alignment has been done and that the pulser, light source and detector are all suitably

timed for fine adjustment work.

The KD*P material used in the cells has a dielectric constant of approximately 80 at the electrical drive frequencies

which are up to around 500MHz. This results in a speed of propagation in the cell of around 7 times slower than the
light. In addition as we are looking for transitions of the order of 700ps or less the separation of the cells in space will
affect the required timing of each cell.

Summary of Contents for Pulse Chopper System

Page 1: ...shops South Moreton Didcot Oxon OX11 9AG U K 10th July 1999 Unit 9 Hall Farm Workshops South Moreton Didcot Oxon OX11 9AG U K Kentech Instruments Ltd Pulse Chopper System J99 10th July 1999 PLEASE REA...

Page 2: ...capacity is small careless use could result in electric shock It is assumed that this highly specialised equipment will only be used by qualified personnel The manufacturers and suppliers accept no re...

Page 3: ...ING THE TIMING OF THE CHANNELS 7 6 0 MOUNTING THE CELLS 9 6 1 PHYSICAL MOUNTING 9 6 2 ELECTRICAL MOUNTING 9 6 3 DC CONNECTIONS 9 Figure Captions Figure 1 The front panel 5 Figure 2 A plot of DC transm...

Page 4: ...ontrol controls the pulse voltage in approximately 20 volt steps from 210 volts to 470 volts All four channels are adjusted together This is effectively the voltage into a 50 load By mismatching the e...

Page 5: ...amplitude polarised at 90 to each other These travel at slightly different speeds in the excited medium and emerge with a relative phase shift of 2 In the unexcited medium there is no relative change...

Page 6: ...s easily obtained from the transmission obtained when the voltage is removed to the transmission when the voltage is applied Note that some power supplies will not discharge the cells when they are tu...

Page 7: ...ming 5 Once all the timings are noted use one in approximately the centre of the range as a reference and adjust all the others so that they are the same see section 5 3 Note the unit supplied is set...

Page 8: ...ot Oxon OX11 9AG U K 10th July 1999 Figure 4 Locating the timing adjusters View with bottom cover removed There are four white adjusters The adjuster at the top of thispicture closesttothe bottom pane...

Page 9: ...ght improve the risetime or not The further apart the cells are mounted the greater the timing for the second cell will have to be adjusted as it will have to be driven slightly later It is possible t...

Page 10: ...s Ltd Unit 9 Hall Farm Workshops South Moreton Didcot Oxon OX11 9AG U K 10th July 1999 1 11 12 13 14 15 16 17 18 19 2 3 4 5 6 7 8 9 10 Figure 5 Output pulse shape one of four channels 100voltsperdivis...

Page 11: ...Hall Farm Workshops South Moreton Didcot Oxon OX11 9AG U K 10th July 1999 Figure 6 The head connected for testing Figure 7 Gating off and looking for good extinction Tested with a Kentech Gated Optica...

Page 12: ...uth Moreton Didcot Oxon OX11 9AG U K 10th July 1999 Figure 8 Testing for long term extinction Showing that reasonable extinction persists to 30ns Figure 9 The gate off fall time Measured with a Kentec...

Page 13: ...13 Kentech Instruments Ltd Unit 9 Hall Farm Workshops South Moreton Didcot Oxon OX11 9AG U K 10th July 1999 Figure 10 The gate off fall time is measured at 728ps...

Reviews: