background image

Kentech Instruments Ltd.

Notes on the use of

CPS3-S module

s/n J19xxxxx

Last Modified 11-2-20

PLEASE READ THIS MANUAL CAREFULLY BEFORE USING THE UNIT

Kentech Instruments Ltd.,

Isis Building, Howbery Park, Wallingford, Oxfordshire, OX10 8BD, U.K.

Tel: +44 (0) 1491 82 1601

E-mail [email protected]    Web www.kentech.co.uk 

Summary of Contents for CPS3-S

Page 1: ...dule s n J19xxxxx Last Modified 11 2 20 PLEASE READ THIS MANUAL CAREFULLY BEFORE USING THE UNIT Kentech Instruments Ltd Isis Building Howbery Park Wallingford Oxfordshire OX10 8BD U K Tel 44 0 1491 82...

Page 2: ...2 20 Contents 1 DECLARATION OF CONFORMITY 3 2 DISCLAIMER 4 3 EMC CAUTION 4 4 ABBREVIATIONS 5 5 CAUTION 5 6 INTRODUCTION 5 7 OPERATION OF THE PULSER 6 7 1 GENERAL OPERATION 6 7 2 CONNECTIONS 6 7 2 1 F...

Page 3: ...3 23 EEC Low Voltage Directive 89 336 EEC Electromagnetic Compatibility Directive 93 68 EEC CE Marking Directive The following harmonised standards have been applied BSEN55011EmissionsSpecification Gr...

Page 4: ...her equipment in its immediate environment It is therefore suitable for use only in a laboratory or a sealed electromagnetic environment unless it is used in a system that has been verified by the sys...

Page 5: ...es incurred in the operation of this equipment Please read the manual before applying power 6 INTRODUCTION Our range of solid state pulsers ASG CPS HMPS and PBG series allows very high voltage fast ri...

Page 6: ...y should then be used If the unit is operated at less than about 11 9 volts the output amplitude may be reduced and the trigger delay increased Also note that the unit is fitted with a 13 volt zener d...

Page 7: ...FRONT PANEL CONNECTIONS This is an N type jack It is important to keep the insulator clean Light lubrication of the thread will stop metal particles from being generated 7 2 2 REAR PANEL CONNECTIONS...

Page 8: ...E TOLERANCE SHEET REMOVE ALL BURRS mm A3 FINISH QTY PER SET DIMS SCALE PAPER 0030 0256 Special CPS3 for Lutronic Corp USED ON ASM 0256 012 MAIN TITLE Kentech Instruments Ltd Isis Building Howbery Park...

Page 9: ...5 5 3 1 BS EN 22768 mK 2 of 3 MATERIAL SPEC SURFACE TEXTURE TOLERANCE SHEET REMOVE ALL BURRS mm A3 FINISH QTY PER SET DIMS SCALE PAPER USED ON ASM 0256 012 MAIN TITLE Kentech Instruments Ltd Isis Bui...

Page 10: ...s BNCtriggerinputtomainoutput Maximum repetition rate 30Hz Power supply 12 Volts DC 0 4 A Maximum power 7W Outputs Pulse output N type 3 6kV pulse Inputs Trigger input BNC jack 5V into 50 5ns rise tim...

Page 11: ...11 Kentech Instruments Ltd Isis Building Howbery Park Wallingford Oxfordshire OX10 8BD U K Last Modified 11 2 20 Figure 1 Pulse shape 4ns...

Page 12: ...12 Kentech Instruments Ltd Isis Building Howbery Park Wallingford Oxfordshire OX10 8BD U K Last Modified 11 2 20 Figure 2 Jitter 5ps SD...