7
Electronic Thermometer
3000
AD/ADA
bio-Tech mode (option Configuration menu)
1. To enter Bio-Tech mode, the thermometer must be in sleep mode (probe securely in place with blank screen).
2. Press and hold the Site and °C/°F buttons at the same time for 4 seconds. A “beep” is heard and a dash scrolls in the display. The
software version of the device will be shown in the alpha-numeric section of the display
3. The thermometer is now in Bio-Tech mode. To navigate the Bio-Tech mode menu, follow the chart below.
4. The Pulse Timer button is used to move forward through the different configuration options. °C/°F button is used to change an
option configuration.
5. To exit Bio-Tech mode and resume normal operation, press the Site and °C/°F buttons at the same time and hold for 1 full second.
Last settings are saved.
6. Bio-Tech mode also exits automatically after 20 seconds of inactivity. The last settings are saved.